John Bunn JB0112-110 User Manual

John Bunn JB0112-110 User Manual

Nebulizer compressor

Advertisement

Quick Links

Neb-u-Lite
Nebulizer Compressor
Item JB0112-110
User Manual
Read this manual before operating the nebulizer.
Save this manual for future reference.
Federal Law (USA) restricts this device to sale by or on
the order of a physician. This device should not be used
unless the operator has been instructed by a qualified
healthcare professional.
LX2
®
JB0112-110-INS-LAB-RevC14

Advertisement

Table of Contents
loading

Summary of Contents for John Bunn JB0112-110

  • Page 1 Neb-u-Lite ® Nebulizer Compressor Item JB0112-110 User Manual Read this manual before operating the nebulizer. Save this manual for future reference. Federal Law (USA) restricts this device to sale by or on the order of a physician. This device should not be used unless the operator has been instructed by a qualified healthcare professional.
  • Page 2: Table Of Contents

    Graham-Field, John Bunn and Neb-u-Lite are registered trademarks of GF Health Products, Inc. © 2007, GF Health Products, Inc. JB0112-110-INS-LAB-RevC14...
  • Page 3: Important Safeguards

    3. Do not place or store device where it can fall or be pulled into a tub or sink. 4. Do not place device in or drop into water or other liquid. 5. Do not reach for a device that has fallen into water. Unplug it immediately. JB0112-110-INS-LAB-RevC14...
  • Page 4 11. Do not use this device in the presence of flammable anesthesia or where oxygen or nitrous oxide is being administered. This device has no AP or APG protection. 12. Connect this device to a properly grounded outlet only. JB0112-110-INS-LAB-RevC14...
  • Page 5 20. Notice for California Customers- California Proposition 65 WARNING: This product contains a chemical known to the State of California to cause cancer and reproductive or developmental harm. SAVE THESE INSTRUCTIONS JB0112-110-INS-LAB-RevC14...
  • Page 6: Introduction

    INTRODUCTION This user manual contains important information and safety precautions for the John Bunn Neb-u-Lite LX2 Nebulizer Compressor. Before using your Nebulizer Compressor, please read and understand this entire user manual. Take special note of all safety precautions marked DANGER and WARNING. Save this user manual for future reference.
  • Page 7: Product Description

    Please note nebulizer features and components, labeled in the illustration below. T adapter storage lid reservoir tube mouthpiece air outlet connector storage compartment filter cap ON/OFF switch nebulizer cup connector tubing Neb-u-Lite LX2 Nebulizer Compressor Info: Neb-u-Lite LX2 replacement parts are available JB0112-110-INS-LAB-RevC14...
  • Page 8 Proceed to step 9. install mask on cap 7. Insert mouthpiece into large end of T adapter T adapter. mouthpiece assemble mouthpiece and T adapter JB0112-110-INS-LAB-RevC14...
  • Page 9 10. Connect the tubing's other end to the compressor air outlet connector. WARNING: Before use, check that all parts are securely connected. Use of an improperly assembled nebulizer could diminish or prevent adequate delivery of medication and the effectiveness of the treatment. JB0112-110-INS-LAB-RevC14...
  • Page 10: Operation

    6. When the treatment is complete, move the ON/OFF switch to the OFF position to turn off the compressor. 7. Unplug the compressor from power outlet. 8. Disconnect nebulizer from tubing. 9. Clean nebulizer as instructed in MAINTENANCE/ Cleaning the Nebulizer Cup. JB0112-110-INS-LAB-RevC14...
  • Page 11: Safety Overload

    2. If the overload protector shuts off the motor frequently, you may have an unstable voltage situation. 3. If the device shuts down and cannot restart, it may need to be replaced. Call your Graham-Field authorized distributor immediately. JB0112-110-INS-LAB-RevC14...
  • Page 12: Maintenance

    2. Once daily, after washing as above, soak nebulizer components in a fresh solution consisting of one part warm water and one part white vinegar for 30 minutes followed by a warm water rinse. Allow to air dry. Store in a clean, dry location. JB0112-110-INS-LAB-RevC14...
  • Page 13: Replacing The Filter

    NOTICE: Do not reuse the filter. Do not operate compressor without a filter. Compressor damage can result. WARNING: The disposable nebulizer kit is for single patient use only. Storage Keep the device and accessories dry. Avoid exposure to direct sunlight. JB0112-110-INS-LAB-RevC14...
  • Page 14: Troubleshooting

    Check tubing for blockage or air leakage at connection to Neb-u-Lite compressor or nebulizer cup; replace as needed. Check that the nebulizer cup is properly assembled and not damaged. If there is any damage, replace as needed. Check if filter is dirty; replace filter as needed. JB0112-110-INS-LAB-RevC14...
  • Page 15: Specifications

    Temperature Operation: 50°F to 104°F (10°C to 40°C) Storage: 5°F to 122°F (-15°C to 50°C) Transport: 50°F to 104°F (-15°C to 70°C) Humidity Operation: 10% to 90% RH non-condensing Storage: 10% to 90% RH non-condensing Transport: 10% to 90% RH non-condensing JB0112-110-INS-LAB-RevC14...
  • Page 16: Limited Warranty

    GF; 6) Any labor or shipping charges incurred in the replacement part installation or repair; 7) Costs and expenses of regular maintenance and cleaning; and 8) Representations and warranties made by any person or entity other than JB0112-110-INS-LAB-RevC14...
  • Page 17 Failure to abide by those regulations may result in a denial of the freight claim. GF will assist you in filing the freight claim. 3) Claims for any short shipment must be made within thirty (30) days of the invoice date. JB0112-110-INS-LAB-RevC14...
  • Page 18 Manufactured for: GF Health Products, Inc. Made in China...