1 Introduction

1.1 Before You Start

Before using qLabs
package inserts for all relevant consumables carefully.

1.2 Intended Use

®
The qLabs
monitoring system, which consists of qLabs
®
and qLabs
Test Strips family, intended for use by healthcare professionals
to perform coagulation tests in a Point of Care environment. The qLabs
monitoring system uses fresh capillary whole blood or fresh venous whole
blood.

1.3 Test Principle

®
The qLabs
ElectroMeter automatically detects the insertion of a qLabs
test strip and heats the strip to a preset operating temperature. After a
drop of blood is applied to the strip, the sample channel automatically
carries the blood sample into the reaction zone through capillary action,
where the blood mixes with pre-printed reagent and starts coagulation
process. As the blood coagulates, conductivity across the test strip
changes. The system measures this electrical change and calculates the
development of a clot. The values are then displayed.

1.4 Packaging

®
The qLabs
ElectroMeter is packed individually.
Upon receiving the package, please open and remove the packaging
materials. Place the qLabs
the power adapter.
Do not keep the qLabs
source, or near an area with a strong magnetic field.
®
ElectroMeter, read the User's Manual as well as the
®
ElectroMeter on a flat surface and connect
®
ElectroMeter in direct sunlight, near a high heat
1
®
ElectroMeter
®
®

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the qLabs ElectroMeter Q-3 Pro and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents