16. Specified of technology

Items
Specifies
Power
Powered by adapter (Model: BJE1M-0060-N600/DA-60I24)
Input: 100~240VAC, 50~60Hz, 2A max
Output: 24V DC, 2.5A
Environment
Temperature
conditions
Humidity
Atmospheric pressure
range
Altitude
Protection class
IP 21, Class Ⅱ
Mode of operation
Continuous operation
Maximum single fault
Device will shut down in the presence of a single fault if the steady state
steady pressure
pressure exceeds: 40 cmH
Sound
Sound pressure level
23
Operate: +5℃ ~35℃ (+41℉ ~95 ℉) ,Non-condensing
Transport: -20℃ ~70℃(-4 ℉ ~158 ℉)
Storage: -20℃ ~70℃(-4 ℉ ~158 ℉)
Operate: relative humidity 15%~90%, Non-
condensing
Transport: relative humidity 15%~90%
Storage: relative humidity 15%~90%
700hPa~1060hPa
≤ 3000m
O
2
Sound pressure level measured according to ISO
80601-2-70:2015 (CPAP mode)
BreathCare PAP ≤ 32dB(A)
Sound power level
Physical properties
Dimensions
(length*width*height)
Weight
Air tube
Maximum volume of
water tank
Materials of water tank PC, Injection molded plastic, stainless steel and
Air outlet
Temperature
Maximum temperature
of heater plate
Cut-out
Maximum gas
temperature
Air filter
Material: Polyester non-woven fiber
Average arrestance: ≥ 85%for~2.5 micron dust
Therapy pressure
4~20 hPa (adjustable, step is 0.5hPa) or 4~20 cmH
Initial pressure
4~20 hPa (adjustable, step is 0.5hPa) or 4~20 cmH
Sound power level measured according to ISO
80601-2-70:2015 (CPAP mode)
BreathCare PAP ≤ 40dB(A)
272.5mm * 181mm * 93mm or 10.73'' * 7.13'' * 3.7''
About 1419 g
Plastic hose,about 1.8m
260±10 mL
silicone seal
22 mm (complies with ISO 5356-1:2015)
55℃ (131 ℉ )(±4℃ )
110℃(if it is damaged, return to manufacturer)
≤ 41℃
O
2
O
2
24