Important Safety Rules Warning: This product is precision manufactured, finely assembled and wired, so do not disassemble or attempt to repair.All repairs CONTENTS must be carried out by qualified personnel at authorized repair centers. Danger: Reduces the risk of electric shock Important Safety Rules -------------- ------ ----1...
5. Negative pressure regulating range: 0.02MPa~limit negative pressrue Product Features 6. Suction rate: 32L/min±2L/min(760mmHg) (Test under operate the suction equipment 10s at the maximum vacuum setting) Application 7. Liquid storage bottle: 1000mL/pc, 1pc Portable phlegm suction unit is based on similar products at home and 8.
Page 4
Open the holder plug; clean up the valve mouth, and leveling the rubber Power line connection valve clack on the float. The valve clack shall not be warped, bent and Connect the plug with the power source. Turn on the power supply, and the broken, but well connected with the float.
shut off. Stop running Emergency measures in the process of application Turn off the aspirator switch, and pull the power plug out of the socket to (1) Quickly loosen the negative pressure regulating knob to release the shut off the power supply. negative pressure if the suction catheter is blocked by strong phlegm and mucus, and start suction again after changing the suction tube.
Maintenance Problem Probable reasons Solution Remark Negative 1)Overflow device 1)After shut-off, turn the 1)Empty the It is recommended to have the suction tube suctioned small amount of holder timely pressure > shut-off regulating valve loose clean water for cleaning up the inner wall before switching off the aspirator. 0.04 MPa, 2)Tube blockage counterclockwise to release...
2. Suction catheter(F8, F12): 1 pc respectively of child & adult by maintaining a minimum distance between portable and mobile RF. 3. Power cord: 1 pc 3. Communications Equipment (transmitters) and the 7E-B5 Portable 4. User's manual: 1 pc Suction.
Page 8
Power magnetic fields should Guidance and manufacturer's declaration – electromagnetic immunity frequency be at levels (50/ 60 Hz) 7E-B5 is intended for use in the electromagnetic environment specified characteristic of a 3 A/m 3 A/m 3 A/m magnetic field below. The customer or the user should assure that it is used in such...
Page 9
If the objects, and people. measured field strength in the location in which the 7E-B5 is used exceeds the applicable RF compliance level above, the 7E-B5 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the aspirators.