BYK-Gardner GmbH Table of Contents Table of Contents 1 Introduction ..........................1.1 Intended Use............................1.2 For Your Safety............................ 1.3 Declarations ............................1.3.1 EU Declaration of Conformity ....................1.3.2 FCC Declaration (USA) ......................1.3.3 IC Declaration (Canada) ......................1.3.4 Japanese Declaration ....................... 1.4 Disclaimer.............................
BYK-Gardner GmbH Introduction | 1 1 Introduction Dear customer, thank you for having decided for a BYK-Gardner product. BYK-Gardner is committed to providing you with quality products and services. We offer complete system solutions to solve your problems in areas of color, appear- ance and physical properties.
1 | Introduction BYK-Gardner GmbH 1.1 Intended Use The spectro2profiler is designed to measure color, gloss, 2D reflectivity and 3D structure on different surfaces. Via the integrated camera the cur- rent spot on the surface can be viewed on the display – allowing exact posi- tioning.
BYK-Gardner GmbH Introduction | 1 1.2 For Your Safety CAUTION Familiarization with Safety Instructions is necessary Absence of knowledge of safety instructions threatens your health and can damage the instrument. Read the safety instructions before you use the in- strument the first time. The safety instructions are part of the delivery content.
1 | Introduction BYK-Gardner GmbH 1.3 Declarations 1.3.1 EU Declaration of Conformity BYK-Gardner GmbH Lausitzer Strasse 8 D-82538 Geretsried declare, that this instrument complies with the requirements of the follow- ing EU directives: • 2014/30/EU - Electromagnetic Compatibility • 2014/35/EU - Low Voltage •...
BYK-Gardner GmbH Introduction | 1 1.3.3 IC Declaration (Canada) This equipment contains a radio module with IC ID 4523A-SN8000. This Class A digital apparatus complies with Canadian ICES-003. This device complies with Industry Canada’s license-exempt RSSs. Operation is subject to the following two conditions: (1) This device may not cause interference;...
Page 10
2 | System Description BYK-Gardner GmbH spectro2profiler 283 025 450 E 2005...
BYK-Gardner GmbH System Description | 2 2 System Description The spectro2profiler is a portable spectrophotometer that measures color, gloss, 2D reflectivity and 3D structure of surfaces. It is operated by the Operate button and the touch-screen display on the top side. The Operate button is used to switch on the instrument and to trigger a measurement.
2 | System Description BYK-Gardner GmbH 2.2 Names and Functions of Parts 2.2.1 Frontside 1 Operate button: 2 Touch-screen display: Press halfway: Touch the icons on the screen with your finger or with a pen • The camera is switched on in order to operate the instru- showing the surface to be ment.
BYK-Gardner GmbH System Description | 2 2.2.2 Backside 1 USB type C plug: • Connected to power-supply: Charge the battery in the instru- ment. • Connected to PC: Charge and transfer data between instrument and computer. 2.2.3 Protective Cap Protective cap: •...
2 | System Description BYK-Gardner GmbH 2.2.4 Power Supply Power supply and adapters: • Use to charge the battery in the instrument. • Use the adapter matching your current wall socket. 2.2.5 USB-Cables USB cables: • Type A/C: Connect instrument with power supply for charging and with PC for charging and data transfer •...
BYK-Gardner GmbH System Description | 2 2.2.6 Checking Tiles In the case following checking tiles are placed: • White standard: Used to calibrate the instrument. • Green standard: Used to test color measurement. • Black standard: Used to test gloss measurement. Consult “Test and Calibration [} 24]“...
2 | System Description BYK-Gardner GmbH The conclusion is: To measure just color is not enough – color and topogra- phy of a sample is to be measured in order to be able to compare it to a standard. The spectro2profiler is the combination of different measure- ment techniques in one instrument: •...
BYK-Gardner GmbH System Description | 2 2.3.3 2D Reflectivity Measurement To overcome the disadvantage mentioned above the spectro2profiler uses a new approach for objective measurement: The illumination takes place in-line; the reflections are measured with a 2D camera. The in-line setup eliminates cast shadows. The advantage is: No shadow zones, no directionality in the measurement results.
Page 18
2 | System Description BYK-Gardner GmbH Summary The spectro2profiler is the combination of different measurement tech- niques in one instrument: • Classic color measurement • Classic gloss and new 2D reflectivity measurement • New 3D topography measurement 1 45/0 color measurement: 2 60°...
BYK-Gardner GmbH Getting Started | 3 3 Getting Started Please observe following notes when you put the instrument into opera- tion: • Assemble the system, consisting of instrument, USB cable, power supply or PC and software. Consult “System Diagram [} 19]” for more infor- mation.
3 | Getting Started BYK-Gardner GmbH 3.2 Charging Procedures The instrument provides two ways for charging: • Standard Charge via USB Type A/C • Fast Charge via USB Type C/C 3.2.1 Standard Charge via USB Type A/C The power supply is connected to the wall socket. Connected the instrument via the USB cable type A/C to the power supply.
BYK-Gardner GmbH Getting Started | 3 3.3 Using the Main Menu The screen below shows the icons that can be displayed on the main menu. Difference Absolute Compare standard and sample. Take absolute measurements Results are saved automatically. without compare. Results are saved automatically.
3 | Getting Started BYK-Gardner GmbH 3.4 Setting Parameters Go to Configuration > Measurement Parameters. A respective list with selectable parameters appears. Consult “Measurement Parameters [} 45]“ for more information. Choose the required parameter and confirm by clicking on the checkmark in the upper right corner.
BYK-Gardner GmbH Getting Started | 3 It is available in two different software packages: “smart-lab” and “smart- process”. The standard delivery includes two licenses for the selected soft- ware package. 3.5.1 Download and Installation Download the ZIP file from: • https://www.byk-instruments.com/spectro2profiler Save the file into a new folder and extract the complete archive.
3 | Getting Started BYK-Gardner GmbH 3.6 Test and Calibration The spectro2profiler has a long-term stable calibration. This can be moni- tored using the calibration standard included in the delivery. If you measure the standard and the values are out of the specification, the instrument will ask you for calibration using this standard.
BYK-Gardner GmbH Perform Measurements | 4 4 Perform Measurements The instrument provides various types of measurement functions: • Difference Measurement: In this mode the results are saved automati- cally. – “Autostandard [} 26]”: Measure sample and search for nearest standard. – “Standard Measurement [} 29]”: Measure sample and save it as new standard.
4 | Perform Measurements BYK-Gardner GmbH In the final state the menu provides following commands • Sample (Rename) • Testserie (Rename) • Delete last measurement • Delete sample • Delete test series • End test series NOTICE Renaming is not supported. You have to give proper name before first measurement.
BYK-Gardner GmbH Perform Measurements | 4 Click on item “Autostandard“. The measurement screen “Standard” is displayed. Measure your current sample or check zone just one times to search for standard. Following scenarios are supported: • Nearest standard found • Some standards close •...
4 | Perform Measurements BYK-Gardner GmbH Measure sample or check zone <n> times and proceed with the next sample or check zone. 4.1.2 Some Standards Close If more than one matching standard is found in instrument, the screen “Some standards close” is displayed. Select the best matching standard from the list and continue measuring.
BYK-Gardner GmbH Perform Measurements | 4 Select the best matching standard from the list and continue measuring. 4.2 Standard Measurement Using this function you can create new standards directly from the sample currently measured. Click on icon “Difference“. The list with existing standards is displayed. Click on the “+”...
Page 30
4 | Perform Measurements BYK-Gardner GmbH Enter a proper name, click on the checkmark symbol to accept and continue measuring. Scroll down to see the measurement details. spectro2profiler 283 025 450 E 2005...
BYK-Gardner GmbH Perform Measurements | 4 Click the arrow symbol to continue with next sample or select “End test series” from context menu. 4.3 Difference Measurement Using this function you can compare standard and sample(s). Click on icon “Difference“. The list with existing standards is displayed. Perform measurement: •...
Page 32
4 | Perform Measurements BYK-Gardner GmbH Scroll down to see data table and statistics. Click the arrow symbol to continue with next sample or select “End test series” from context menu. spectro2profiler 283 025 450 E 2005...
BYK-Gardner GmbH Perform Measurements | 4 4.4 Absolute Measurement Using this function you can take measurements without compare to stan- dard. Click on icon “Absolute“. The measurement continues at your last test se- ries. Perform measurement: • Click on “Measure” or press “Operate” button. •...
4 | Perform Measurements BYK-Gardner GmbH 4.5 Perform Quick Check Using this function you can compare a standard with one or more samples without saving for a quick evaluation. Click on icon “Quick check“. The measurement starts with the standard. Perform measurement: •...
BYK-Gardner GmbH Perform Measurements | 4 4.6 Opacity Measurement With this function you can measure the hiding power of your using - for ex- ample - our byko-charts, see https://www.byk-instruments.com/c/p-5916. Note that the option “Opacity“ has to be activated in the Color indices, see: “Setting Parameters [} 22]”.
4 | Perform Measurements BYK-Gardner GmbH 4.7 Organizer Usage Organize the Measurement Process It is possible to predefine measurement sequences with the software “smart-process”. The measurement can be accelerated via so-called “orga- nizers”. The “Organizer” icon appears after downloading at least one organizer from “smart-process”...
BYK-Gardner GmbH Browse Measurements | 5 5 Browse Measurements Using this function you can view measured data and / or delete existing standards or test series. Click on icon Browse. The list with all types of measurements is displayed. You can browse through the list in order to view or delete the data. 5.1 View Measurements In the list of measurement types open the object for which you want to view the measured data –...
Page 38
5 | Browse Measurements BYK-Gardner GmbH Open the test series to view the measurement data. Scroll down to view values and statistics. spectro2profiler 283 025 450 E 2005...
BYK-Gardner GmbH Browse Measurements | 5 Browse through the different sample measured. 5.2 Delete Complete Standard You can delete complete standards with all their measured data. Select the object(s) to be deleted in the list. Click on the Trashcan icon. A confirmation dialog is displayed. 283 025 450 E 2005 spectro2profiler...
5 | Browse Measurements BYK-Gardner GmbH The selected objects are deleted. 5.3 Delete Measurement Data Do not select the standard in the list but click the standard to open it. The test serie is listed. Select the test serie to be deleted and click on the Trashcan icon. A confir- mation dialog is displayed.
Page 41
BYK-Gardner GmbH Browse Measurements | 5 The selected object is deleted. 283 025 450 E 2005 spectro2profiler...
BYK-Gardner GmbH Configuration | 6 6 Configuration With this function you can configure the system according to your needs. For example: You can set time and language or choose various measure- ment parameters. Click on icon Configuration. The configuration screen is displayed. There are different types provided to change the configuration of the system.
6 | Configuration BYK-Gardner GmbH 1 Back: Go back to previous 5 Active: Option is activated. screen. Click left to deactivate. 2 Cancel: Go back to previous 6 Wheel: Tap an entry in the screen without saving. menu and move it to the left or to the right.
BYK-Gardner GmbH Configuration | 6 4 Date / Time: Set system time, 10 Factory Reset time zone and daylight saving Reset instrument to factory set- time here tings here. Consult “Factory Re- set [} 50]” for more informa- tion. 5 Color Scheme 11 Protect configuration: If acti- vated a password is required to Adjust screen brightness to day...
6 | Configuration BYK-Gardner GmbH Default is none. 5 Gloss 11 Always continue last test se- ries Turn gloss measurement on or off. No new test series is created; latest series is opened automat- ically. 6 ΔGloss 12 Measurement Screen Turn difference measurement Select data to be displayed on for gloss on or off.
BYK-Gardner GmbH Configuration | 6 Following values are possible (in minutes): 1, 2, 5, 10, 15, 30. 6.2.4 Test and Calibration The system delivery comprises a calibration tile and test tiles for color and gloss. These tiles can be used to check if your instrument is measuring cor- rectly.
Page 48
6 | Configuration BYK-Gardner GmbH Place the instrument on the tile and select Configuration > Calibration. NOTICE Once you have measured your test tiles you can save them as standard. This way you can use “Difference Measurement [} 31]” later on to check the dif- ference between test tile and instrument.
Page 49
BYK-Gardner GmbH Configuration | 6 You can check date of last calibration and serial numbers for your calibra- tion and test tiles. Scroll down to see all details. 6.2.4.2 Autodiagnosis You will be informed by the instrument, when a diagnosis is required. Se- lect menu option Autodiagnosis.
6 | Configuration BYK-Gardner GmbH 6.2.4.4 Color-Test Place instrument on green standard and perform following steps: 1. Select menu option Test Standard Color. 2. The instrument gives you instructions and automatically guides you through the test. 3. Follow the instructions of the instrument. The green standard is measured.
BYK-Gardner GmbH Configuration | 6 After fallback all your personal configuration and measurement data in the device will be lost. The instrument is set back to original state. 6.2.6 Protect Configuration You can protect the current configuration of your instrument via password against accidental or intentional changes.
Page 52
6 | Configuration BYK-Gardner GmbH If the option is activated the Configuration menu can only be accessed from the dashboard by entering the password. Note down the password on a secure place – if you do not remember, you will have to contact your local BYK-Gardner certified service center to get the master password.
BYK-Gardner GmbH Troubleshooting | 7 7 Troubleshooting Problem/Error message Solution Error! Please clean Clean white tile and repeat white White Tile or call calibration. customer service. If white calibration fails again, con- tact service. Calibration Perform calibration using the white invalid.
Page 54
8 | Technical Data BYK-Gardner GmbH spectro2profiler 283 025 450 E 2005...
BYK-Gardner GmbH Technical Data | 8 8 Technical Data Color Geometry 45°c:0° Aperture Size Diameter 25 mm Spectral Range 400–700 nm, 10 nm resolution Repeatability 0,01 ∆E94 (10 readings on white) Reproducibility 0,1 ∆E94 (average of 12 BCRA tiles) Color Systems CIELab/Ch, Lab(h), XYZ, Yxy Color Differences ∆E*, ∆E(h), ∆E94, ∆ECMC, ∆E99, ∆E2000...
Page 56
8 | Technical Data BYK-Gardner GmbH General Data Memory 3000 samples with images 10 000 samples without images Languages English, German, French, Italian, Spanish, Russian, Japanese, Chinese Dimensions (LxWxH) 150 x 240 x 155 mm (5,9 × 9,5 × 6,1 in) Weight 1530 g (3,37 lbs) Interface...
BYK-Gardner GmbH Service Points | 9 9 Service Points BYK-Gardner global service centers with ISO / IEC 17025 accredited laboratories Headquarter Ger- Headquarter USA Headquarter PTE many c/o BYK-Gardner USA c/o BYK USA dba Paul N. c/o BYK-Gardner GmbH Gardner 9104 Guilford Rd., Co- Lausitzer Strasse 8, lumbia, MD 21046, USA...