Pasco Scientific ES-9054B Instruction Manual And Experiment Manual

Electrometer

Advertisement

Quick Links

Instruction Manual and
Experiment Guide for
the PASCO scientific
Model ES-9054B
ELECTROMETER
© 1987 PASCO scientific
Model ES-9054B ELECTROMETER
scientific
INPUT
OUTPUT
GND
FUNCTION
B1 B2
D
3
C
10
V
O
30
L
T
100
S
ZERO
ADJUST
POWER
ON
OFF
ZERO
PUSH
LOCK
TO
ZERO
012-03175E
3/93
$7.50

Advertisement

Table of Contents
loading

Summary of Contents for Pasco Scientific ES-9054B

  • Page 1 012-03175E Instruction Manual and 3/93 Experiment Guide for the PASCO scientific Model ES-9054B ELECTROMETER FUNCTION B1 B2 Model ES-9054B ELECTROMETER ZERO ADJUST POWER scientific INPUT OUTPUT ZERO PUSH LOCK ZERO © 1987 PASCO scientific $7.50...
  • Page 3: Table Of Contents

    012-03175E Electrometer Table of Contents Section Page Copyright and Warranty ..................ii Equipment Return ...................ii Introduction ..................... 1 Operation ......................1 Setup ......................2 Measuring Charge ..................3 Measuring Voltage ..................4 Measuring Current ..................4 Using a Projection Meter ................5 Maintenace ......................
  • Page 4: Equipment Return

    AUTHORIZATION FROM PASCO. Limited Warranty When returning equipment for repair, the units must be PASCO scientific warrants the product to be free from packed properly. Carriers will not accept responsibility defects in materials and workmanship for a period of for damage caused by improper packing. To be certain one year from the date of shipment to the customer.
  • Page 5: Introduction

    012-03175E Introduction The Model ES-9054B Electrometer is an infinite With these features, you’ll find that your electrostatics impedance (10 ý) voltmeter that can be used for demonstrations and labs are easier to perform and, with direct measurements of voltage, and indirect measure- quantitative data, are more informative as well.
  • Page 6: Setup

    (adjust with power off) Select voltage range (3, 10, 30, or 100 volts full scale) or FUNCTION test batteries B1 B2 (B1 and B2) Model ES-9054B ELECTROMETER ZERO ADJUST Zero meter (zero POWER scientific switch should be in INPUT OUTPUT...
  • Page 7: Measuring Charge

    012-03175E Measuring Charge Measuring Charge by Induction Under most conditions, the best way to measure charge Charge measurement with the electrometer is indirect, is by induction, using a proof plane and a Faraday ice but simple. It is based on the relationship Q = CV, pail such as those included with PASCO’s Demonstra- where tion Electrostatics System.
  • Page 8: Measuring Voltage

    Using a Projection Meter will need to repeat the above procedure to measure the new capacitance. The OUTPUT connectors on the ES-9054B Electrom- eter provide an output signal for driving a projection NOTE: The capacitance of the Electrometer ä...
  • Page 9: Using A Projection Meter

    012-03175E PASCO’s Model ES-9065 Projection Meter can be Turn the ZERO ADJUST knob until the meter on the Electrometer reads as close to full scale as used with an overhead projector to display the Elec- possible (either postive or negative). trometer readings for the whole class.
  • Page 10: Mos-Fet Replacement

    Transistor). You can order a replacement from covered in the preceding sections, we strongly suggest PASCO scientific (Part # 422-001) . When ordering a you return it to PASCO scientific for repair. replacement part, be sure to include the model number of the Electrometer.
  • Page 11: Schematic

    012-03175E SCHEMATIC Model ES-9054B Electrometer (Drawing #956-02859)
  • Page 12: Parts List

    012-03175E Parts List Reference Description PASCO Part # R1,2,5,13, Resistor, 10 ký, 1/4 W, 5% 113-103 20,21 Resistor, 1 ký, 1/4 W, 5% 113-102 Resistor, 13 ký, 1/4 W, 5%, C.C. 113-133 R6,24 Resistor, 470 ý, 1/4 W, 5% 113-471 Potentiometer, 10 ký, 1/4 W, 30% 140-006 Resistor, 330 ý, 1/4 W, 5%...
  • Page 13: Technical Support

    012-03175E Electrometer Technical Support Feedback Contacting Technical Support If you have any comments about the product or manual, Before you call the PASCO Technical Support staff, it please let us know. If you have any suggestions on would be helpful to prepare the following information: alternate experiments or find a problem in the manual, If your problem is with the PASCO apparatus, note: please tell us.