WPI MityFlex Instruction Manual

Peristaltic pump

Advertisement

Quick Links

INSTRUCTION MANUAL
MityFlex
Peristaltic Pump
Serial No._____________________
WORLD
PRECISION
INSTRUMENTS
Instrumenting scientific ideas
www.wpiinc.com
051316

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MityFlex and is the answer not in the manual?

Questions and answers

Summary of Contents for WPI MityFlex

  • Page 1 WORLD PRECISION INSTRUMENTS Instrumenting scientific ideas INSTRUCTION MANUAL MityFlex Peristaltic Pump Serial No._____________________ www.wpiinc.com 051316...
  • Page 3: Table Of Contents

    MityFlex CONTENTS ABOUT THIS MANUAL ........................1 INTRODUCTION ..........................1 SPECIFICATIONS ..........................2 OPERATING INSTRUCTIONS ......................3 Installation ..........................3 Roller/Tubing Selection ......................3 Variable Speed ........................... 3 Tubing Installation ........................4 Alternate Tubing Installation Procedure ................ 4 Fuse Installation ........................5 Motor Brush Replacement .....................
  • Page 4 World Precision Instruments...
  • Page 5: About This Manual

    NOTES and TIPS contain helpful information. INTRODUCTION MITYFLEX™ variable speed drive is powered by DC gearmotors. A specially designed electronic speed control system provides smooth acceleration control over the entire flow range. The three position on/off switch allows the user to maintain speed setting when unit is turned off or reversed.
  • Page 6: Specifications

    FUSE: Located in the inlet receptacle. Uses 5 X 20mm - 1.0 amp. (See page 4 for replacement procedure.) DC MOTOR BRUSHES: Replaceable (See page 4 for replacement procedure.) WEIGHT: 13 lb Included with MityFlex are four short lengths of Norprene tubing to fit into the roller ™ ” , ”...
  • Page 7: Operating Instructions

    MityFlex OPERATING INSTRUCTIONS Installation POWER REQUIREMENTS: Voltage and frequency of power supply must be the same as shown on unit. WIRING CONNECTIONS: All wiring and electrical connections must comply with local and national electrical codes. MOUNTING: Place on any flat surface. Ensure that space is provided at the back of the case for air circulation through ventilation holes.
  • Page 8: Tubing Installation

    Tubing Installation Inspect all tubing regularly and replace if any sign of deterioration occurs. Always wear safety glasses and protective clothing when working with chemicals. Disconnect power to the pump. Disconnect the suction and discharge tubing from the pump tubing. Remove four thumb screws and pump cover.
  • Page 9: Fuse Installation

    MityFlex Fuse Installation Disconnect power to pump. Find the power socket on the back of the case. Pull the power cord out. Slide the drawer below the receptacle out toward you. Remove old fuse and replace with new fuse. Push drawer in until it locks in place.
  • Page 10: Declaration Of Conformity

    DECLARATION OF CONFORMITY World Precision Instruments...
  • Page 11: Warranty

    WPI shall not be liable for any damage to data or property that may be caused directly or indirectly by use of this product.
  • Page 12 International Trade Center, 175 Sarasota Center Blvd., Sarasota FL 34240-9258 Tel: 941-371-1003 • Fax: 941-377-5428 • E-mail: sales@wpiinc.com 1 Hunting Gate, Hitchin, Hertfordshire SG4 0TJ Tel: 44 (0)1462 424700 • Fax: 44 (0)1462 424701 • E-mail: wpiuk@wpi-europe.com Germany Zossener Str. 55, 10961 Berlin Tel: 030-6188845 •...