Technical Specifications - Prestige medical DT-32 Instructions Manual

Table of Contents

Advertisement

TECHNICAL SPECIFICATIONS

• Dimensions: 140 mm (L) x 43 mm (W) x 47 mm (H)
• Weight: Approximately 93 g (including batteries)
• Range for measurement of body temperature:
93.2°F (34°C) ~ 108°F (42.2°C)
• Range for measurement of the temperature of objects:
32.18°F (0.1°C) ~ 211.8°F (99.9°C)
• Resolution: 0.1°C
• Accuracy under laboratory reference conditions:
Body Mode Temperature Accuracy:
±0.2°C in the range from 35°C ~ 42°C
±0.3°C in the range < 35°C and > 42°C
Object Mode Temperature Accuracy:
±1.0°C in the range from 0.1°C ~ 99.9°C
• Measurement Distance: Max. 2 inches (5 cm)
• Display: LCD
• Memory: 30 Measured Values
• Backlight: Blue
• Operating Conditions:
Temperature range: 59°F (15°C) ~ 104°F (40°C)
Humidity (relative): up to 95% (no condensation)
• Storage and Transport:
Temperature Range: -4°F (-20°C) ~ 131°F (55°C)
Humidity (relative): up to 95% (no condensation)
• Automatic Shut-Off:
60 seconds after a temperature measurement
30 seconds in memory mode
10 seconds after an error message or after a devia-
tion in ambient temperature from the defined
temperature range
• Voltage/Batteries: 1.5V x 2 AAA (LR03)
• Life of Batteries: At Least 1,000 Measurements
• Anticipated Service Life of the Device: 10 Years
• Protection Against Penetration by Water or Solid
Materials: IP 22