  • Page 2: Unit Description

    UNIT DESCRIPTION 1. O/I Button 2. Display 3. Start Button 4. Sensor 5. Mode Switch 6. Battery Cover   DISPLAY DESCRIPTION 1. Temperature reading 2. Object Temperature Icon 3. Body Temperature Mode Icon 4. Low Battery Warning Icon 5. Ready for Measurement Icon 6.
  • Page 3: Advantages Of This Thermometer

    ADvANTAgES Of ThIS ThERmOmETER measurement in a matter of seconds The innovative infrared technology allows measurement without touching the object for safe and sanitary measurements within seconds. multiple Uses (wide range measurement) This thermometer offers a wide measurement range feature from 0 °C to 100.0 °C (32.0 °F to 212.0 °F), meaning the unit can be used to measure body temperature or it can be used to measure surface temperature of the following:...
  • Page 4: Important Safety Instructions

    ImPORTANT SAfETY INSTRUCTIONS This instrument may only be used for the purposes described in this booklet. The manufacturer cannot be held liable for damage caused by incorrect application. • Never immerse in water or other liquids. For cleaning please follow the instructions in the Cleaning and Disinfecting section.
  • Page 5: How This Thermometer Measures Temperature

    hOW ThIS ThERmOmETER mEASURES TEmPERATURE This thermometer measures infrared energy radiated from the forehead or objects. This is collected through the lens and converted to a temperature value. Pointing at the area above the eyebrow will provide the greatest accuracy. CONTROL DISPLAYS AND SYmbOLS •...
  • Page 6: Operating Notes

    mEASURINg IN bODY mODE 1. Press the O/I button. The display is activated to show all segments for 2 seconds. 2. The last measurement reading will automati- cally be shown on the display for 2 seconds with the “M” icon at the lower right corner.  ...
  • Page 7 • Don’t take a measurement immediately after nursing a baby. • Don’t use in high humidity environments. • Patients should not drink, eat, or exercise before/while taking the measurement. • Don’t move the measurement device from the measuring area before hearing the termination beep.
  • Page 8 ChANgINg ThE UNIT Of mEASUREmENT fROm CELSIUS TO fAhRENhEIT 1. Make sure the unit is turned off. 2. Press and hold the START button for more than 5 seconds until the current unit of measurement is displayed. 3. Press the START button again to switch between °C (Celsius) and °F (Fahrenheit).
  • Page 9: Error Messages

    ERROR mESSAgES • Measured temperature too high: Displays “H” when   measured temperature is higher than 42.2 °C (108.0 °F) in body mode or 100 °C (212 °F) in object mode. • Measured temperature   too low: Displays “L” when measured temperature is lower than 34.0 °C (93.2 °F) in body mode or 0 °C (32 °F) in object...
  • Page 10: Cleaning And Disinfecting

    CLEANINg AND DISINfECTINg Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer casing and the measuring probe. Ensure that no liquid enters the interior of the thermometer. Never use abrasive cleaning agents, thinners or benzene for cleaning and never immerse the instrument in water or other cleaning liquids.
  • Page 11: Technical Specifications

    TEChNICAL SPECIfICATIONS Type: Non Contact Thermometer [DT-30] measurement range: • Body Mode: 34.0-42.2 °C (93.2-108.0 °F) • Object Mode: 0-100.0 °C (32-212.0 °F) Resolution: 0.1 °C measurement accuracy: Laboratory: ±0.3 °C, 36.0 ~ 39.0 °C (±0.4 °F, 96.8 ~ 102.2 °F)
  • Page 12: Reference To Standards

    Automatic Switch-off: Approx. 1 minute after last measurement has been taken. battery: CR2032 Battery (X1) 3V - at least 1000 measurements Dimensions: 112±0.3mm × 30±0.2mm × 24±0.2 mm Weight: 50±0.5 g (with battery), 47±0.5 (w/o battery) Reference to standards: ASTM E1965-98; IEC 60601-1; IEC 60601-1-2 (EMC);...