TABLE OF CONTENTS I. SAFETY PRECAUTIONS………………………………………………………………………3 II. INTRODUCTION………………………………………………………………………………4 III. PRECAUTIONS BEFORE USE……………………………………………………………4 IV. OPERATING PRINCIPLE……………………………………………………………………5 THE DIFFERENT METHODS OF TEMPERATURE MEASUREMENT…………………5 NORMAL TEMPERATURES ACCORDING TO MEASUREMENT METHOD…………6 NORMAL TEMPERATURES ACCORDING TO AGE………………………………………7 PRACTICAL CONSIDERATIONS WHEN TAKING A TEMPERATURE………………8 HOW TO TAKE A TEMPERATURE……………………………………………………………8 ADDITIONAL CONSIDERATIONS BEFORE USE……………………………………9 V.
- Do not expose the instrument to sunlight or to water. - Do not drop the instrument. - Should a problem occur with your DT-29 Infrared Forehead Thermometer, please contact your dealer or Prestige Medical. Do not attempt to repair this instrument yourself.
(1). However, as with other types of thermometers, it is essential to use the DT-29 properly in order to obtain reliable and stable results. You are therefore advised to read this instruction manual and observe the safety precautions carefully before use.
All objects, solid, liquid or gas, emit energy by radiation. The intensity of this energy depends on the temperature of the object. The DT-29 Infrared Forehead Thermometer is able to measure the temperature of a person by the heat energy the person emits.
Oral thermometry Oral temperature is easily influenced by recent ingestion of food or drinks and by breathing through the mouth. To measure oral temperature, the mouth must remain closed and the tongue lowered for three to four minutes, a difficult task for young children to accomplish.
The DT-29 No-Touch Thermometer has been designed to produce an instant forehead temperature representing core temperature reading without touching the skin. Since the body temperature is permanent and regular at the forehead, it allows precise measurement of the temperature.
PRACTICAL CONSIDERATIONS WHEN TAKING A TEMPERATURE - In order to ensure that precise and accurate temperature measurements are obtained, it is essential that each user has received adequate information on and training in the temperature measurement technique when using this device. - It is essential to remember that although procedures such as taking a patient’s temperature may be simple they must not be trivialized.
- Allow a 1 minute interval between two measurements. - Each time there is a significant change in the ambient temperature due to a change in the environment, let the DT-29 acclimate to this new ambient temperature for at least 15 minutes before using it.
6. Data displayed in Celsius or Fahrenheit. 7. Automatic shut-off (energy saver). 8. Small, convenient, easy to use. Additional function: DT-29 Infrared Forehead Thermometer can also be used to measure the temperature of a baby-bottle or bath,( by using the Surface Temp function) or room temperature (by using the Room function).
VIII. SETTING AND FUNCTION OF MENU CONFIGURATION AND FUNCTION OF MENU 1. Setting the mode a. Press Mode button, and the screen will display: Body…°F b. Press again and the screen will display: Surface Temp…°F c. Press again and the screen will display: Room…°F Note: The instrument default is set to BODY mode 2.
Page 13
5. Recalibration of device via the F4 MENU The DT-29 calibration is factory-set. Depending on various skin types and thickness, there may be temperature differences. The F4 function will recalibrate and if you see a difference in the reading, you must recalibrate the unit.
10. Accuracy: ± 0.6° F (0.3° C) 11. Measuring distance: 1.2 – 2 in (3 cm – 5 cm) 12. Automatic turn-off: 30 sec. DT-29 : Precision Precision is not guaranteed outside temperature readings below 90°F or above 109°F (32°C to 42.9°C).
X. WARNINGS - The protective glass over the lens is the most important and fragile part of the thermometer, please handle with care. - Clean the glass with cotton swab, dampened with water or 70% alcohol solution. - Use alkaline batteries whenever possible. DO NOT use rechargeable batteries. - Remove the batteries when Instrument is not being used for an extended period of time.
XII. TROUBLESHOOTING If you have one of the following problems while using your DT-29 No-Touch Thermometer, please refer to this breakdown service guide to help resolve the problem. If the problem persists please contact our Technical Service Department. THE SCREEN DISPLAYS A BODY TEMPERATURE LESS THAN 89.6°F ( 32°C) To take a body temperature the function must be on Body mode.
Page 17
THE SCREEN DISPLAYS THE MESSAGE : LO When using the DT-29 No-Touch Thermometer, the message Lo can appear on the screen. The temperature measured is under the measuring range selected, either less than 90°F (32°C) in Body Mode or less than 32°F (0°C) in Surface Temp Mode.
Need help?
Do you have a question about the DT-29 and is the answer not in the manual?
Questions and answers