Measurement Out of Specified Conditions:
• The thermometer automatically measures the ambient
temperature in body temperature mode and object
mode. If the ambient temperature is above or below the
defined temperature range, a warning is displayed and
the temperature measurement is not carried out.
• If the ambient room
temperature is more than
104°F (40°C), the warning
notice "Er.H" appears and
there are 3 short acoustic
signals. The thermometer
will then switch off after 10
seconds.
• If the ambient room tem-
perature is too low in body
temperature mode [less than
59°F (15°C)] or is too low in
object mode [less than 41°F
(5°C)], the warning notice
"Er.L" appears and there are 3 short acoustic signals. The
thermometer will then switch off after 10 seconds.
If this occurs, make sure to use the thermometer under
the specified environmental conditions. Move the
device to an environment with the specified conditions
and wait 30 minutes before using it.

STORAGE AND MAINTENANCE

• In order to ensure appropriate performance and long
service life, the thermometer should be stored in a dry
location between -4°F (-20°C) ~ 131°F (55°C).
• Avoid storing in direct sunlight, and protect it from
shocks and vibrations. Do not store in a wet or high
humidity location, or in the vicinity of heat sources.
• Remove the batteries from the device if you plan to
store it for an extended period of time.