Page 1
Dynamic Platform Model 008PL USER GUIDE 900 Asbury Ave • Norfolk, Virginia 23513 • USA Tel: 800.617.1177 • 757.855.2765 • Fax: 757.857.0523 WWW.CIRSINC.COM Tissue Simulation & Phantom Technology...
Page 2
Overview Model 008PL The CIRS Dynamic Platform provides an economical, user-friendly solution Features for the complex tasks associated with tumor motion and patient position- ing in radiation therapy. • Move any phantom with The platform is made from stiff, low-density plastics. The device enables...
Page 3
Unpacking Instructions Remove cables from case and set aside. Remove foam insert from case and set Before you open the case check the aside. three Drop ‘N’ Tell indicators on the right side of the case. Drop ‘N’ Tell shipping damage indicator shows when a case has been dropped in transit and contains potential dam- aged goods.
Page 4
Assembly Procedure WARNING: FOLLOW THE CABLE CONNECTION STEPS AS THEY ARE PRESENTED IN THIS USER GUIDE. CONNECTING THE CABLES WITH THE CONTROLLER “POWER ON” CAN SERIOUSLY DAMAGE THE PHANTOM’S ELECTRONICS. This will secure standard Plastic Water Other phantoms may require user to drill ®...
Page 5
Adapter” folder found on the provided USB drive or download tion is finished, plug the “USB-to-Network Adapter” into the zipped folder from the CIRS website. Unzip to a known your PC’s USB port and acknowledge the Windows installa- location and follow instructions from “Instructions.pdf”...
Page 6
Select Internet Protocol Version 4 (TCP/IPv4) and click on prop- In the internet Protocol Version 4 (TCP/IPv4) properties window, erties change from default “Obtain an IP address automatically” to “Use the following IP address:” Enter “192.168.0.101” as the IP address and “255.255.255.000” as the Subnet mask. Note: If an IP address conflict occurs because IP address 192.168.0.101 is already assigned to another Net work Adapter, the user can try any other IP address between...
Page 7
512 MB RAM, 2 MB of available disk space If the end-user is offline during use of the phantom, it is recom- mended that a copy of the CIRS Motion Control User Manual is INTRODUCTION downloaded and saved. Once a copy of the manual is saved in a...
Page 8
Safety GENERAL SAFETY NOTICES WARNING: TO REDUCE THE RISK OF FIRE, ELECTRIC The following general safety notices supplement the specific warn- ings and cautions appearing elsewhere in this manual. They are rec- SHOCK, OR INJURY WHEN USING THE MO- ommended precautions that must be understood and applied during TION CONTROLLER, FOLLOW THESE BASIC operation and maintenance of the equipment covered herein.
Page 9
5.4 kg (12 lbs.) Motion Accuracy: ±0.1 mm Cycle Time: 1 - ∞ (adjusted based on amplitude) Editable built-in waveforms: sin(t), 1-2cos4(t), 1-2cos6(t), sawtooth, sharkfin CIRS Motion Control Software System Requirements Windows XP or later ® Pentium 3 or equivalent ®...
Page 10
CIRS not to comply with documented order specifications. You must 24 Months Platform return the product to CIRS within 30 calendar days of the issu- ance of the RMA. All returns should be packed in the original cases and or packaging and must include any accessories, manuals and There are no warranties, expressed or implied, including without documentation that shipped with the product.
Need help?
Do you have a question about the 008PL and is the answer not in the manual?
Questions and answers