Download Print this page

Maintenance - M-system M2VF Manual

Signal transmitter

Advertisement

TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the con-
nection diagram on the side of the unit.
When an input resistor is provided with the module, attach
it together with input wiring to the input screw terminals.
+
SIGNAL
R
SOURCE
*Input shunt resistor attached for current input.
The M2VF, by its fast-response feature, is not designed to eliminate
noise present in the input signal. Use a shielded twisted-pair cable for
preventing noise entering through the input wiring.
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
13 – 14 with a multimeter.
3) Input: Check that the input signal is within 0 – 100% of
the full-scale.
4) Output: Check that the load resistance meets the de-
scribed specifications.
EXTERNAL DIMENSIONS
21.5 (.85)
10 (.39)
M-SYSTEM WARRANTY
M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System's sole liability, and purchaser's exclusive
remedies, under this warranty are, at M-System's option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this
warranty. To submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original
sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: info@m-system.co.jp
SEN TRONIC
AG
Produkte, Support und Service
+
1
9
*
OUTPUT
4
12
13
U(+)
POWER
14
V(–)
unit: mm (inch)
84 (3.31)
114 (4.49)
• When mounting, no extra space is needed between units.
Rugghölzli 2
CH - 5453 Busslingen
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■ HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply 0% input and adjust output to 0%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).

MAINTENANCE

Regular calibration procedure is explained below:
■ CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the "ADJUST-
MENT PROCEDURE" explained earlier.
6 (.23)
DIN RAIL
4
1
35mm wide
8
5
2–4.2x5
(.17x.20)
MTG HOLE
6 (.24) deep
12
9
8–M3
SCREW
14 13
15 (.59)
[4 (.16)]
23 (.91)
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
M2VF
■ TERMINAL ASSIGNMENTS
INPUT RESISTOR
(model: REM2)
4
1
8
5
12
9
14 13
Input shunt resistor attached
for current input.
EM-504 Rev.7
P. 2 / 2
mailbox@sentronic.com
www.sentronic.com

Advertisement

loading

This manual is also suitable for:

Mini-m seriesM2vf-1M2vf-12M2vf-3M2vf-34