Instructions for Use
Note: Wavelength range can be especially useful to physicians, such as those performing
photodynamic therapy.
Note: A functional tester cannot be utilized to assess the accuracy of Current Health SpO2
monitoring.
18.4.3 Respiratory Rate
Sensing Method
Measurement Range
Resolution
Accuracy
Data Update Period
Apnea Detection
Chest Movements
6rpm – 60rpm
1rpm
Supine Arm at Side: 1rpm
Supine Arm Above Head: 2rpm
Seated at 90°: 2rpm
Seated at 45°: 1rpm
Standing: 3rpm
Prone: 2rpm
Lying on Side: 1rpm
28 seconds
No
www.currenthealth.com
200002/1
Page 60 of 61