EMC Guidance
The ME EQUIPMENT or ME SYSTEM is suitable for home
healthcare environments.
Warning Don't near active HF surgical equipment and the
RF shielded room of an ME system for magnetic
resonance imaging, where the intensity of EM
disturbances is high.
Warning Use of this equipment adjacent to or stacked with
other equipment should be avoided because it
could result in improper operation. If such use is
necessary, this equipment and the other equip-
ment should be observed to verify that they are
operating normally.
Warning Use of accessories, transducers and cables other
than those specified or provided by the manufac-
turer of this equipment could result in increased
electromagnetic emissions or decreased electro-
magnetic immunity of this equipment and result in
improper operation.
Warning Portable RF communications equipment (including
peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm
(12 inches) to any part of the equipment BC 87,
including cables specified by the manufacturer.
Otherwise, degradation of the performance of this
equipment could result.
Technical description
1. 1 all necessary instructions for maintaining BASIC SAFE-
TY and ESSENTIAL PERFORMANCE with regard to elec-
tromagnetic disturbances for the excepted service life.
2. Guidance and manufacturer's declaration -electromag-
netic emissions and Immunity
Table 1
Guidance and manufacturer's declaration -
electromagnetic emissions
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
232
Compliance
Group 1
Class B
Not applicable
Not applicable