Fuses have to be replaced by types as rated on the nameplate on the rear of the unit. Personal safety must be given the highest priority! hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany +49 40 537122-0 ...
Page 4
Sollte ein Ersatz der Netzsicherungen erforderlich sein, so ist sicherzustellen, dass nur Sicherungen der angegebenen Nennstromstärke und Nennspannung als Ersatz verwendet werden. Der Personensicherheit ist höchste Priorität einzuräumen! hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany +49 40 537122-0 ...
For HA2B5-S output voltages of -2000V to +2000V at load currents of up to ±5mA and >7mA provided. The signal gain is 200, the input voltage range is -10V…+10V.
(180° phase shift in respect to amplifier 1). This mode can be used e. g. to differentially drive two deflection plates or to drive a floating load with up to 8000V hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany ...
Achievable slew rate: SR = I [V/s]. If the average positive or negative output current exceeds 5mA (HA2B5-S) or 3mA (HA3B3-S) the amplifier shuts down with Overload. If the maximum slew rate is exceeded, distortions occur in the output signals. In principle, the amplifier should thus only be controlled by input signals that satisfy the achievable slew rate for a given load.
The device must be disconnected from the power supply before carrying out any operations on the test setup or before touching the output terminals. hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany +49 40 537122-0 +49 40 537122-99 info@hivolt.de...
- Signal ground and high voltage ground are connected to chassis ground / protective earth and the earth connector (M4 nut). - The appliance conforms to protection category I. It must only be used on mains power sockets with a ground connection. Amplifier modules HA2B5-S / HA3B3-S Parameter Conditions Supply voltage, V ±10%...
- Operation - Storage and Transportation - Depending on the ambient temperature, modulation amplitude and load capacitance, the maximum output power may need to be derated. hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany +49 40 537122-0 ...
This symbol on the output terminals warns of the risk of electric shock. Temperature Compensation To avoid condensation within the device, it should be allowed to reach the room temperature. Please unpack the product at least two hours prior to power-up. hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany ...
15. The Interlock LEDs will light up continuously. 16. Press the SDN Reset push button. 17. The interlock status remains unchanged and the high voltage cannot be switched on. 18. Turn off the supply voltage. hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany ...
Overload, Overtemperature) is no longer pending. It is recommended to operate the unit with shutdown latches enabled and actively reset any shutdown condition via the SDN Reset push button or the SDN signal on X1. hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany ...
5.11 Internal Connectors Internal connector X1 of HA2B5-S / HA3B3-S plug-in modules (DIN41612, Type F) Pins Signal Function n.c. n.c. n.c. n.c. D10 B10 Z10 n.c. D12 B12 Z12 GND D14 B14 Z14 n.c. D16 B16 Z16 n.c. HVON_BP HV enable input; used for interlock or general high voltage ON;...
Caution: disconnect the unit from the power supply before removing any modules or opening any covers. Further regular maintenance is not required. Cleaning If necessary, wipe the device with a slightly damp cloth. Do not use abrasive detergents or solvents. hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg Germany +49 40 537122-0 ...
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements Classification: Group 1, Class B Manufacturer: hivolt.de GmbH & Co. KG Oehleckerring 40 D-22419 Hamburg 2021-03-09 Wulf Müller Managing Director hivolt.de GmbH & Co. KG...