Operations Manual
Item: H2020 and H2024 (-E)
I. Specifications:
Speed:
H2020
Fixed 24 RPM (
20RPM for 230V, 50Hz)
H2024
Adjustable from 1 to 70rpm
Temp. Range: Ambient +5 to 60C
Temp. Fluctuation: 0.9C
Max Capacity: 6 x 50ml / 16 x 15ml / 44 x 1.5-2.0ml
Dimensions: (wxdxh) 29.5 x 21.6 x 22.9cm (11.6x8.5x9 in.)
Weight: 9lbs / 4kg
Electrical: 120V, 60HZ or 230V, 50Hz, 2A 250W
Operating environment: 20°C to 45°C
Warranty: 2 years
II. Product Set-Up:
Place the rotator on a clean and stable surface.
(Ensure that there is at least 3 inches of open space
V1.0
both in front of and behind the instrument.) Plug
the instrument into an outlet of the appropriate
voltage.
III. Tube Holder Installation / Exchange:
The instrument includes an assortment of tube
holders and posts.
ALWAYS attach the tube holder load tubes outside
of the incubation chamber. Once fully loaded, then
load the posts with tube holders onto the rotating
axis. Loading tubes and tube holder within the
incubation chamber may result in the dropping of
tubes and spillage on the heating surface.
1.5/2.0, 5, 7, 10 and 15ml Tubes:
Attach the tube holders (above) to the included post
with two arms. Press holders on firmly to prevent
them from coming loose during use. Once pressed
on firmly, attach the post to the rotator by inserting
the post into the left side of the rotator first and then
into the main body (right side). Once inserted, rotate
the post until it is fully installed, see below:
To adjust the angle of which the tubes rotate, simply
rotate the position of the tube holders by hand until
the desired rotational angle is achieved.
0.5 and 50ml Tubes
Attach
the
holder
directly
rotator
(without
Insert the holder into the
left side of the rotator first
and then into the main
body (right side).
inserted, rotate the post
until it is fully installed (similar to diagram above).
To rotate tubes horizontally (rolling/hybridization
motion), first attach the 2 small posts to the center of
the large tube holder. (One post has a small flat that
can be inserted to correspond to the flat located in
the center of the tube holder). Thread the 2 posts
together by turning the post (without flat) clockwise
until fully tightener.
Then insert posts into the
rotator as described above.
To load test-tubes, press them firmly into the
appropriate holder. ALWAYS ensure a balanced load
across both sides of the rotational axis.
mixing 15 and 50ml tubes, always load a similar
quantity of tubes on each side of the rotational axis.)
large
tube
to
the
posts.)
Once
(When

Advertisement

loading

Summary of Contents for Benchmark H2020

  • Page 1 Place the rotator on a clean and stable surface. (Ensure that there is at least 3 inches of open space the desired rotational angle is achieved. Item: H2020 and H2024 (-E) V1.0 both in front of and behind the instrument.) Plug the instrument into an outlet of the appropriate 0.5 and 50ml Tubes...