Configuration with Monitor Replacement
System
Multistar KK with
display 2000 60Hz
AXIOM Artis Bypass (!)

Jumper Configuration (Hardware)

NOTE
Jumpers for Monitor Size
Behind the front plane, on the digital
controller, there are two jumpers:
X11 => configuration of the monitor
size. Incorrect configuration leads to
the error: EEPROM (in this case, re-
position the jumper).
- 1 installed => 44cm (17")
- not installed => 54cm (21")
X8 => Sensitivity of the ambient light
sensor (factory setting, 2-3 middle).
Video Terminal Resistor / Cable Compensator
Remove the rear wall of the housing
Remove the 2 screws on the rear wall, take
the housing off towards the back.
Remove the top cover panel.
Siemens AG
Medical Solutions
Norm Used
Norm 7 (2045
lines/120Hz/1:1)
Norm 3 (1125
lines/60Hz/1:1)
If one of the jumper positions described below is changed, the
monitor must be disconnected from line power!
RA52-060.061.01.07.02
07.02
Blocked Norm (black background)
Norm 8 (1023 lines/120Hz/4:3)
n.a.
Fig. 2:
Monitor size
Fig. 3:
Page 15 of 24
TD SD 21
15
0