Replacing the Firmware
5-

FW Replacement

4Replacing the Firmware
Unplug the power connector.
The FW is located on the digital controller.
The digital controller is located behind the
front control panel.
Unlatch the front control panel (on both
sides) using the unlatching tool that is includ-
ed in the shipment, and flip it down.
Disconnect the ribbon cable from the digital
controller to the monitor.
Remove the metal shielding (2 screws).
Replace the FW.
Installation is in the reverse order of the
steps above.
NOTE
Siemens AG
Medical Solutions
Insert the screw inserted in plastic as follows:
Turn the screw counterclockwise until it snaps in place, only then
turn it clockwise to tighten it.
RA52-060.061.01.07.02
07.02
Fig. 1:
Page 11 of 24
TD SD 21
11
0
SIMOMED HM