Transmitter Unit - Ancillary Equipment; Recorders; Range Indication; Sensor Unit - Access To Terminals - ABB 7835 User Manual

Hydrazine monitor
Table of Contents

Advertisement

7835
Hydrazine Monitor
4.4 Transmitter Unit – Ancillary Equipment

4.4.1 Recorders

The choice of 2 different isolated recorder output signals
enables the monitor to be used with a wide variety of recording
and data processing equipment. A 2-pen recorder is necessary;
pen 1 indicating the hydrazine concentration and pen 2 the
monitor range.

4.4.2 Range Indication

The remote range indication relays can be used in different
arrangements to suit the requirements of the installation. The
relays can, for example, be wired directly into the PLC or data
logger but, if a recorder is used, a method of indicating the
range set is required.
Suitable range indication recorder input can be provided using a
resistor network that consist of 4,
network as shown in the two examples in Fig. 4.4. A recorder
with suitable voltage and resistance inputs can be provided by
ABB. The recorder provides 60 % scale deflection for Range 1
and 70 % scale deflection for Range 2.
Ensure other arrangements are designed to suit
requirements of the system. Ensure that all external equipment
is set up and working according to the instructions supplied.
30R
To Chart
Recorder
0 to 100 Ω Input
0 V
30R
To Chart
Recorder
10 V DC External Supply
0 to 100 mV Input
Fig. 4.4 Resistor Network for Remote Range Indication
(0 to 100 Ω and 0 to 100 mV Input Options)
10
1
/
watt resistors. Connect the
4
39R
Range 1
Range 2
30R
39R
Range 1
Range 2
30R
10K
4.5 Sensor Unit – Access to Terminals
Warning. Isolate the transmitter from power supplies
before removing the cover.
Referring to Fig. 4.5:
1. Unlock the door
2. Remove all 4 screws
C
plate
.
the
Fig. 4.5 Sensor Unit – Access to Terminals
4 Electrical Installation
A
and open fully.
B
and remove the terminal cover
A
C
B
IM/7835–EN Rev. P