5

Setup the WorkHorse Mariner ADCP

Figure 9
illustrates how to connect the WorkHorse Mariner cables on your
workbench. You will need a container of water large enough to submerge
the WorkHorse's transducer head into during testing (two to three inches of
water is sufficient). Testing the WorkHorse out of water may cause some
tests to fail but causes no harm to the WorkHorse.
INPUT POWER
DC 20-60 VOLT
DC 12 VOLT
AC 98-264 VOLT
DECK BOX (REAR VIEW)
FROM CHANNEL 1
(J20 RS-232 or J19 RS-422)
TO COMPUTER
TO
ENSEMBLE-OUT
DEVICE
Figure 9.
5.1

Serial Communication

The standard communications settings for WorkHorse Mariners is RS-422,
9600-baud, no parity, 8 data bits and 1 stop bit. The Deck Box converts
P/N 957-6169-00 (April 2005)
• The deck box is not designed to withstand water. Use cau-
tion when using on decks in wet conditions.
• Avoid using ferro-magnetic materials in the mounting
fixtures or near the Workhorse. Ferro-magnetic materials
affect the compass.
TO ADCP
(CHOOSE ONE INPUT VOLTAGE)
(OPTIONAL)
COMPUTER
WorkHorse Mariner Connections
WorkHorse Mariner ADCP User's Guide
TO J17
I/O CABLE
FROM (J22)
TO GYRO
EXTERNAL RS-232 (J28)
GYROCOMPASS DATA
TO COMPUTER
FROM
NAVAGATION
(OPTIONAL)
DEVICE