Guidance and Manufacturer's Declaration
Below cables information are provided for EMC reference.
Max. cable length,
Cable
Shielded/unshielded
DC Power Line
0.92m
Important information regarding electromagnetic compatibility (EMC)
This electrical medical equipment needs special precautions regarding EMC and should be
put into service according to the EMC information provided in the user manual. The equipment
conforms to this IEC 60601-1-2:2014 standard for both immunity and emissions. Nevertheless,
special precautions need to be observed:
Warning: The equipment is intended used in professional healthcare facility environment ex-
cept for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM
for magnetic resonance imaging, where the intensity of EM DISTURBANCES is high. Essential
Performance of the device includes supplying continuous sound output during a medical
evaluation equal to or below 92.1 dB.
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally.
Warning: The use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Warning: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the 658,
including cables specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Warning: When the AC input voltage is interrupted, the equipment will shut down and if the
power supply restored, it should be recovered by operator manually. This degradation could be
accepted because it will not lead to unacceptable risks and it will not result in the loss of basic
safety or essential performance.
Warning: Accessories and cables provided with this device have not been tested or verified
for use with other medical electrical equipment and should not be considered interchangeable.
EMI Compliance Table (Table 1)
Phenomenon
RF emissions
Harmonic distortion
Voltage fluctuations
and flicker
EMS Compliance Table (Tables 2-4)
Table 2 - Enclosure Port
Phenomenon
Basic EMC Standard
Electrostatic discharge
Radiated RF EM field
Proximity fields from RF wireless
communications equipment
Rated power frequency
magnetic fields
Table 3 – Proximity fields from RF wireless communications equipment
Test Frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
Table 4 – Input a.c. power Port
Phenomenon
Basic EMC Standard
Electrical fast transients/burst
Surges line-to-line
Conducted disturances
induced by RF fields
Voltage dips
Voltage interruptions
Number
classification
Shielded
1 Set
Table 1 - Emission
Compliance
Electromagnetic environment
CISPR 11
Professional healthcare
Group 1, Class B
facility environment
IEC 61000-3-2
Professional healthcare
Class A
facility environment
IEC 61000-3-3
Professional healthcare
Compliance
facility environment
Immunity test levels
Professional healthcare
facility environment
±8 kV contact
IEC 61000-4-2
±2kV, ±4kV, ±8kV, ±15kV air
3V/m
IEC 61000-4-3
80MHz-2.7GHz
80% AM at 1kHz
IEC 61000-4-3
Refer to table 3
30A/m
IEC 61000-4-8
50Hz or 60Hz
Band
Immunity test levels
(MHz)
Professional healthcare
facility environment
380-390
Pulse modulation 18Hz, 27V/m
430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
704-787
800-960
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
1700-1990
2400-2570
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
5100-5800
Immunity test levels
Professional healthcare
facility environment
±2 kV
IEC 61000-4-4
100kHz repetition frequency
±0.5 kV, ±1 kV
IEC 61000-4-5
3V, 0.15MHz-80MHz
IEC 61000-4-6
6V in ISM bands between 0.15MHz
and 80MHz
80%AM at 1kHz
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º
and 315º
0% UT; 1 cycle
IEC 61000-4-11
and
70% UT; 25/30 cycles
Single phase: at 0º
IEC 61000-4-11
0% UT; 250/300 cycles
Cable
DC Power
Instructions
for Use
Warranty
ADC
®
warrants its products against defects in materials and workmanship under
normal use and service as follows:
1. Warranty service extends to the original purchaser from the date of purchase.
2 years - Electronic Chestpiece
Lifetime - Binaurals & Tubing
2. The entire stethoscope is warranted. This warranty is only valid if the product
is registered online at www.adctoday.com.
What Is Covered: Repair, or replacement of parts, and labor.
What Is Not Covered: Transportation charges to ADC. Damages caused by abuse,
misuse, accident, or negligence, incidental, special, or consequential
damages. Batteries or damage from old batteries is not covered by the warranty.
Some states do not allow the exclusion or limitation of incidental, special, or
consequential damages, so this limitation may not apply to you.
To Obtain Warranty Service: Send item postage paid to ADC, Attn: Service Dept.,
55 Commerce Dr., Hauppauge, NY 11788. Please include your name and
address, daytime phone no., proof of purchase, and a brief note explaining the problem.
Implied Warranty: Any implied warranty shall be limited in duration to the terms of
this warranty and in no case beyond the original selling price (except where prohibited
by law). This warranty gives you specific legal rights and you may have other rights
which vary from state to state.
©2019 ADC
®
. All rights reserved. No one is permitted to reproduce or duplicate in any form, this manual or
any part thereof without the expressed written permission of ADC. ADC assumes no responsibility for any in-
jury to anyone, or for any illegal or improper use of the product that may result from failure to use the product
in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
ADC
®
55 Commerce Drive
Hauppauge, NY 11788
IB p/n 93-658-00 rev 2.5
ADC
Adscope
®
Electronic Stethoscope
To register your product visit us at
www.adctoday.com/register
FOR QUESTIONS, COMMENTS,
OR SUGGESTIONS CALL TOLL FREE:
1-800-ADC-2670
OR VISIT
www.adctoday.com/feedback
This manual is available online in a variety of
languages, follow the links for language options.
www.adctoday.com/care
658
®
®
Inspected, assembled and
packaged in the U.S.A.
Chestpiece made in China
Headset made in Taiwan
tel: 631-273-9600
1-800-232-2670
fax: 631-273-9659
www.adctoday.com
Printed in the U.S.A.

Advertisement

loading

Summary of Contents for ADC Adscope 658

  • Page 1 Table 3 – Proximity fields from RF wireless communications equipment any part thereof without the expressed written permission of ADC. ADC assumes no responsibility for any in- jury to anyone, or for any illegal or improper use of the product that may result from failure to use the product...
  • Page 2 You'll quickly appreciate the results, for you now own one of the finest electronic stetho- After the 658 powers on, pressing either the Bell or Diaphragm mode button will change scopes that money can buy. With proper care and maintenance, your Adscope 658 is sure to the mode.

This manual is also suitable for:

Adscope 658bdAdscope 658nAdscope 658dgAdscope 658mrsAdscope 658mcaAdscope 658mcb ... Show all