12. CARE AND CLEANING

To ensure accurate measurements, it is very important to keep the
infrared sensor clean and free of scratches. Fingerprints and dirt will
affect the accuracy of the thermometer. In order to get an accurate
measurement, please clean the infrared sensor. Gently wipe its surface
with a cotton swab slightly moistened with alcohol and immediately
wipe dry with a clean cotton swab.
Use a soft dry cloth to clean the body of the thermometer. Never clean
the thermometer using an abrasive cleanser, thinner or benzene, nor
immerse the thermometer in water or other liquids. After cleaning,
place the thermometer in its protective case. Store it in a clean, dry
place at room temperature. Never expose the thermometer to extreme
temperatures, humidity, direct sunlight or shock.
• Clean the measuring sensor after each use. Use a clean cloth or
cotton bud that can be moistened with 75% alcohol.
• To clean the entire device, please use a soft cloth
slightly moistened with a mild soapy solution. Under no
circumstances may liquid enter the device. Do not use the device
again until it is completely dry.
13. CLINICAL ACCURACY VALIDATION
METHOD
This device is an adjusted mode clinical thermometer the validated
information for clinical accuracy in each adjusted mode are:
GroupA1: ∆cb=-0.01
GroupA2: ∆cb= 0.06
GroupB : ∆cb=-0.01
GroupC : ∆cb=-0.01
∆cb :
CLINICAL BIAS
L
:
LlMITS OF AGREEMENT
A
σr :
CLINICAL REPEATABILITY
14
C, LA=0.18, σr=±0.08
0
C, LA=0.22, σr=±0.08
0
C, LA=0.20, σr=±0.07
0
C, LA=0.18, σr=±0.07
0
C
0
C
0
C
0
C
0
Care And Cleaning