Signal Connection - National Instruments SCXI-1325 Installation Manual

High-voltage terminal block
Hide thumbs Also See for SCXI-1325:
Table of Contents

Advertisement

device to National Instruments for service and repair to ensure that its safety features are
not compromised.
When using the device with high common-mode voltages, you must insulate your signal
wires for the highest input voltage. National Instruments is not liable for any damages or
injuries resulting from inadequate signal wire insulation. Use only 26-14 AWG wire with
a voltage rating of 300 V and 60 °C for signals that may come in contact with 250 V.
When connecting or disconnecting signal lines to the SCXI terminal block screw terminals,
make sure the lines are powered off. Potential differences between the lines and the SCXI
ground create a shock hazard while you connect the lines.
Connections, including power signals to ground and vice versa, that exceed any of the
maximum signal ratings on the SCXI device can create a shock or fire hazard or can
damage any or all of the boards connected to the SCXI chassis, the host computer, and the
SCXI device. National Instruments is not liable for any damages or injuries resulting from
incorrect signal connections.
If high voltages (≥30 V
earth ground wire to the terminal block safety ground solder lug, shown in Figure 1. This
complies with safety agency requirements and protects against electric shock when the
terminal block is not connected to the chassis. To connect the safety earth ground to the
safety ground solder lug, run an earth ground wire in the cable from the signal source to
the terminal block. National Instruments is not liable for any damages or injuries resulting
from inadequate safety earth ground connections.
Do not loosen or re-orient the safety ground solder lug hardware when connecting the
safety ground wire; to do so reduces the safety isolation between the high voltage and
safety ground.

Signal Connection

Note
Refer to the
connecting or disconnecting any signal wires.
© National Instruments Corporation
and 42.4 V
rms
peak
Safety Information
When connecting your signals to the SCXI-1325, follow the labeling on the
SCXI-1325 for the appropriate module, as indicated in Figure 2.
To connect the signal to the terminal block, perform the following steps,
referring to Figures 1 and 2 as necessary:
1.
Unscrew the top cover screws and remove the cover.
2.
Loosen the strain-relief screws and remove the strain-relief bar.
3.
Run the signal wires through the strain-relief opening. You can add
insulation or padding if necessary.
4.
Prepare your signal wire by stripping the insulation no more than
7 mm.
or 60 VDC) are present, you must connect a safety
section before removing equipment covers or
3
SCXI-1325 Installation Guide