 CLAMP MULTIMETER
E N G LI SH
Start Guide
English
PRECAUTIONS FOR USE
This device complies with safety standards IEC-61010-1 and 61010-2-032 for voltages of
1000V in category IV at an altitude OF less than 2000m, indoors, with a degree of pollution
not exceeding 2.
These safety instructions are intended to ensure the safety of persons and proper operation of
the device. If the tester is used other than as specified in this data sheet, the protection
provided by the device may be impaired.
The operator and/or the responsible authority must carefully read and clearly understand
the various precautions to be taken in use.
If you use this instrument other than as specified, the protection it provides may be
compromised, thereby endangering you.
Do not use the instrument in an explosive atmosphere or in the presence of flammable
gases or fumes.
Do not use the instrument on networks of which the voltage or category exceeds those
mentioned.
Do not exceed the rated maximum voltages and currents between terminals or with
respect to earth.
Do not use the instrument if it appears to be damaged, incomplete, or not properly
closed.
Before each use, check the condition of the insulation on the leads, housing, and
accessories. Any element of which the insulation is deteriorated (even partially) must be
set aside for repair or scrapped.
Use leads and accessories rated for voltages and categories at least equal to those of
the instrument. If not, an accessory of a lower category lowers the category of the
combined Clamp + accessory to that of the accessory.
Observe the environmental conditions of use.
Do not modify the instrument and do not replace components with "equivalents". Repairs
and adjustments must be done by approved qualified personnel.
Replace the batteries as soon as the
Disconnect all cords before opening the battery compartment cover.
Use personal protective equipment when conditions require.
Keep your hands away from the unused terminals of the instrument.
When handling the test probes, crocodile clips, and clamp ammeters, keep your fingers
behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the device, we
recommend performing configuration operations only when the device is disconnected
from all dangerous voltages.
F401
Clamp Multimeter F401
symbol appears on the display unit.
Clamp Multimeter F401
You have just acquired an F401 clamp multimeter and we thank you.
For best results from your device :
read this user manual attentively,
observe the precautions for its use.
Meanings of the symbols used on the device
Danger. The operator agrees to refer to this data sheet whenever
this danger symbol is encountered.
Application or withdrawal authorized on uninsulated or bare
conductors at dangerous voltages.
1.5 V battery.
The CE marking indicates compliance with European directives.
Double insulation or reinforced insulation.
Selective sorting of wastes for the recycling of electrical and
electronic equipment within the European Union.
In conformity with directive DEEE 2002/96/EC: this equipment must
not be treated as household waste.
AC – Alternating current.
AC and DC – Alternating and direct current.
Earth.
Risk of electric shock.
Clamp Multimeter F401
MEASUREMENT CATEGORIES
Definitions of the measurement categories :
CAT II: Circuits directly connected to the low-voltage installation.
Example: power supply to household electrical appliances and portable tools.
CAT III: Power supply circuits in the installation of the building.
Example: distribution panel, circuit-breakers, fixed industrial machines or devices.
CAT IV: Circuits supplying the low-voltage installation of the building.
Example: power lines, meters, and protection devices.
1
PRESENTATION
1
2
7
3
4
5
6
English
English
Item
Designation
Jaws with centring marks (see
1
connection principles)
2
Physical guard
3
Switch
4
Function keys
5
Display unit
6
Terminals
7
Trigger

Advertisement

loading

Summary of Contents for Chauvin Arnoux F401

  • Page 1  CLAMP MULTIMETER Clamp Multimeter F401 English F401 You have just acquired an F401 clamp multimeter and we thank you. For best results from your device :  read this user manual attentively,  observe the precautions for its use.
  • Page 2 English Clamp Multimeter F401 Clamp Multimeter F401 English THE ROTARY SWITCH THE DISPLAY UNIT The rotary switch has five positions. To access the , functions, set the rotary Item Function switch to the desired function. Each setting is confirmed by an audible signal. The functions are described in the table below ;...