STORING VALUES INTO MEMORY
There are 5 memory locations for each function. The default values stored in the locations are:
Memory
Type
Location
J, K, C, R, S, N
o
0.0
C
M1
o
(32.0
F)
o
100.0
M2
o
(212.0
o
500.0
M3
o
(932.0
o
750.0
M4
o
(1382.0
o
1000.0
M5
o
(1832.0
Memory
mA
Location
4.00mA
M1
8.00mA
M2
12.00mA
M3
16.00mA
M4
20.00mA
M5
To change the values in memory:
1.
Select the SOURCE mode
2.
Press the SET button to select the memory location to be changed.
3.
Press the arrow buttons to adjust to the new value
4.
Momentarily press the MEM button to store the value. The memory location icon will blink
while the value is being stored..
Type
T
o
0.0
C
o
(32.0
F)
o
C
100.0
C
o
F)
(212.0
F)
o
C
200.0
C
o
F)
(392.0
F)
o
C
300.0
C
o
F)
(572.0
F),
o
C
400.0
C
o
F)
(752.0
F)
%
0.0%
25%
50%
75%
100%
6
Type
T/C mV
E
o
0.0
C
0mV
o
(32.0
F)
o
100.0
C
10.00mV
o
(212.0
F)
o
200.0
C
25.00mV
o
(392.0
F)
o
500.0
C
40.00mV
o
(932.0
F)
o
750.0
C
50.00mV
o
(1382.0
F)
mV
V
0mV
0V
500mV
5V
1000mV
10V
1500mV
15V
2000mV
20V
PRC30-en-EU_V1.3 7/14