Sample Calculation, Buffered Output; Derivation Of Integrated Signal Voltage - GE Bently Nevada 3300/25 Operation Manual

Dual accelerometer input monitor
Table of Contents

Advertisement

3300/25 Dual Accelerometer Input Monitor Operation Manual
21.2.1

Sample Calculation, Buffered Output

Monitor is configured as: BB or CC option 16 (50 mm/sec (peak))
Input vibration level is: 1g (peak)
Input signal frequency is: 100 Hz = 6000 CPM
Transducer scale factor is: 100 mV/g
Step 1. Calculate transducer output voltage.
Vib = 1g (peak)
SF = 100 mV/g
Step 2.
The signal at the buffered output is 0.3091 V (peak) or equivalently 0.2186 V (RMS).
21.2.2

Derivation of Integrated Signal Voltage

The transfer function of the integrator and the gain block is:
Z
f
( Y
) S
=
×
4
Z
i
Where:
Z
is the feedback impedance:
f
Z
is the input impedance:
i
S
=
× i
ω
=
× i
2
×
π
×
138
1
( g
peak
× )
0
1 .
×
29
Vout
=
(.
009097
×
R
×
f
Z
=
f
R
+
f
Z
=
R
I
I
F
0
1 .
=
0
1 .
V
=
Vsig
V
g
.
645
=
. 0
3091
2
100
+ )
1
1
SC
f
1
SC
f
V
(
peak
)