Siemens SIMATIC OP7 Manual page 32

Programme for chambers with mplc, by vicpas
Hide thumbs Also See for SIMATIC OP7:
Table of Contents

Advertisement

10 (°C)
To set segment
10 (% RH)
To set segment
Leave everything switched OFF as humidity is not controlled in this temperature range
SHIFT
ENTER
This page is visualized
• Press
repeatedly
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
Temperature from 95°C to 110°C
Maximum speed
hh : mm : ss
Durat.
0 : 0 : 0
Wait duration
Channel 1
Control
Wait set point
Max speed
:110.00 °C
Set point
: 0.00 °C/min
Grad.
Near set
Humidity - - - - not controlled
Save segment
Save segment
Total segment
Confirm progr
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 11
N1
: OFF
N2
: ON
: ON
: ON
N3
: 1,00 °C
: SAVE
to save the programmed segment
When this page is visualized, if the previ-
: OFF
ously described operations have been
carried out, the segment has been saved
and therefore must not be saved again.
: 0
Move on to the next step.
: OFF
N4
Wait for this number to appear.
31
110
95
To carry out the segment at maximum
speed
10

Hide quick links: