Siemens SIMATIC OP7 Manual page 27

Programme for chambers with mplc, by vicpas
Hide thumbs Also See for SIMATIC OP7:
Table of Contents

Advertisement

5 (°C)
To set segment
5 (% RH)
To set segment
SHIFT
ENTER
This page is visualized.
• Press
repeatedly
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
Temperature 55°C
Duration 2,5h - Maintenance
hh : mm : ss
Durat.
2
: 30
Wait duration
Channel 1
Control
Wait set point
Max speed
: 55.00 °C
Set point
: 0.00 °C/min
Grad.
Near set
Humidity from 95% to 90% and maintenance
Duration 2.5h - Maintenance
Channel 2
Control
Wait set point
Max speed
Set point
: 90.00 RH %
Grad.
: 20,00 RH %/min
Near set
: 3,00 RH %
Water recharge:
Not used:
Not used:
SP. hum. CTRL.
Aux 1:
Aux 2:
Aux 3:
Aux 4:
Save segment
Save segment
Total segment
Confirm progr
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 6
F1
: 0
: OFF
F2
: ON
: ON
: OFF
F3
: 1,00 °C
F4
: ON
: OFF
: OFF
F5
F6
OFF
OFF
OFF
ON
To improve the system performance
(experimentally set)
F7
In this segment the AUX1 must be
ON
switched ON
OFF
OFF
OFF
F9
to save the programmed segment
: SAVE
When this page is visualized, if the previ-
: OFF
ously described operations have been
carried out, the segment has been saved
and therefore must not be saved again.
: 0
Move on to the next step.
: OFF
F10
Wait for this number to appear.
26
30
Because the same temperature value
is maintained
95
Unreal gradient which enables maxi-
mum speed to be obtained
55
5
90
5

Hide quick links: