Technical Data Of Sensor - Elis FLONET FN50 Series Design, Assembly And Service Manual

Electromagnetic flowmeter
Table of Contents

Advertisement

ELIS PLZEŇ a. s.

4.1.9. Technical data of sensor

Dimensions of sensor
Operating pressure
Connection of sensor
Earthing
Flow rate of measured liquid
Temperature of measured liquid
Minimum conductivity of measured liquid
Lining
Electrodes
Ingress protection
Storage temperature
ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722
Design, Assembly and Service Manual
Electromagnetic flowmeter FLONET FN50xx.1
Flangeless DN 25 to DN 200
Flanged DN 15 to DN 800
PN 40 (4,0 MPa) for DN 15 to DN 50
PN 16 (1,6 MPa) for DN 65to DN 200
PN 10 (1,0 MPa) for DN 250to DN 800
Flangeless
Flanged
Earthing electrode
0,05 m/s to 10 m/s (range 1:200)
o
110
C
Up to 150ºC on a request (based on lining type and
when using sensor with prolonged chimney made of
stainless material) – see article 4.1.4
20 μS/cm, based on agreement with manufacturer up
to 5 μS/cm
Soft rubber
Hard rubber
Rubber for potable water
Teflon PTFE
E - CTFE
Stainless steel 1.4571 (316Ti)
Hastelloy C276
Tantalum
Titanium
Platinum-Rhodium (PtRh10)
IP 65
-10ºC to +70ºC, at the max. relative humidity 70 %
Page 17 of 32
Es90716K

Advertisement

Table of Contents
loading

This manual is also suitable for:

Flonet fn50.1

Table of Contents