In the following table, for the example, the temperature and the corresponding voltage current and
power parameter settings are reported for activation and conditioning of the most popular SAES
getter pumps, like the CapaciTorr and the NEXTorr families, without jacket.
NAME
CT D100
CT D200
CT D400-2
CT D50
CT HV 200 35
CT HV 200 63
CT Z100
CT Z200
CT Z400 body
CT Z400 nude
GP50-2F
NXT D100-5
NXT D200-5
NXT D300-5
NXT D500-5
NXT HV 100
NXT HV 200
NXT HV 300
NXT Z100
NXT Z200
NXT Z300
NXT D500-StarCell
ATTENTION: SAES Getters is not responsible for damages due to improper use of the
power supply
M.HIST.0138.23 R1
Activation
Conditioning
Mode
Mode
Vmax
Vmax
9 V
3.5 V
12 V
4 V
16 V
7 V
8 V
3 V
11.3 V
3.3 V
12.6 V
3.7 V
8 V
4 V
11 V
5 V
15 V
5 V
18 V
7 V
25 V
11 V
9 V
4.5 V
12.5 V
6 V
20 V
8 V
28 V
10 V
9 V
2.8 V
12 V
3.5 V
30 V
8 V
9 V
4 V
12 V
5 V
24 V
9 V
28 V
10 V
Page 17 of 28