In the Run screen it is possible to change Vset value without interrupting output supply: this result
can be achieved by keeping pressed MENU/ENTER and acting at the same time on the arrow keys.
The new value will be active as soon as MENU/ENTER is released.
Please note that this value replaces the setting and after the end of current supply the Vset in the
Standby page will be also updated with the new value.
4.2.

Setting Menu

With a long press on MENU/ENTER it is possible to enter in the pages to configure the instrument
parameters.
The pictures show the default parameters of NEG POWER MINI.
The first (read only) page shows the software version, the hardware version and the MAC address.
With the down arrow you can scroll to the second page.
Here MODBUS TCP can be enabled and IP parameters can be changed.
A short pressing on MENU/ENTER activates editing of the first field and moves to the next field,
arrow keys change parameter values. After editing the last parameter (Port) the next pressing on
MENU/ENTER ends editing.
The following parameters are available:
Enable/Disable: Enable/Disable the Modbus over TCP/IP on the Ethernet interface
IP: IP address, the four bytes can be edited separately.
GW: the IP address of the gateway, the four bytes can be edited separately.
Mask: is the IP address net mask encoded as the number of bits corresponding to the network
portion of the IP address (for example 24 corresponds to 255.255.255.0)
Port is the TCP port where communication takes place.