Bio-logic
NavPRO ONE
®
®
QUICK START GUIDE
Select the test ear by pressing either the Right or Left button, or Right + Left for binaural testing. Binaural testing is available when
headphones, insert phones or two ear probes are connected (if stimulating at low intensity levels only).
Make sure the patient is sitting or lying down comfortably.
An impedance check will be performed prior to measurement. If impedance is below 6 k ohm, the start button will become
available. You can configure the settings for the test to autostart if the impedance is lower than 4 k ohm. If impedance is higher
than 6 k ohm but below 12 k ohm, a skip button will become available. An impedance below 6 k ohm is recommended for best
test results.