Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE To check or re-calibrate Tymp ECV Calibration: Connect the red connector from the tymp probe cable to the red jack on the back of tubing (air connector) securely to the pump outlet to the right of the blue connector on the back of the device with a gentle twisting motion until it is finger tight.
Page 3
Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE You will be in “checking“ mode. You can now insert the probe into any of the 3 cavities to check the current volume reading. Please make sure to use the white probe tip as shown below. White (small) probe tip If the readings are fine (standards allow 5% or 0.1 ml, whichever is greater), simply press the Back button.
Page 4
Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE This procedure will repeat for all three cavities, and display a message upon success If for any reason the calibration does not succeed, you will see this message. Possible reasons for calibration fails are: •...