QUICK START GUIDE
OAE audiometry and
tympanometry screening

Advertisement

loading

Summary of Contents for natus Bio-logic AuDX PRO FLEX

  • Page 1 Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE OAE audiometry and tympanometry screening...
  • Page 2: Device Connectors

    Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE Device connectors Headphones or On/Off button Power supply USB cable Probe second transducer Label printer Patient response button Air Pump Start menu Please note that not all Start Menu options will be available depending on the configuration and modules on your device.
  • Page 3 Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE Testing: Tympanometry Connect the probe to the red connector and the air cable to the pump connector. From the start menu, select the Immittance: Tymp+Ref button. Select the test ear by pressing either the Right or Left button. Cartoon mode will show an airplane on the screen during the tympanometry test.
  • Page 4 Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE Distortion Product Otoacoustic Emissions (DPOAE) Screening Select either the Screening or the Standard button from the start menu and connect the probe to the red connector, or use the tympanometry probe. Select the appropriate ear tip size and place it on the probe tip.
  • Page 5 Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE DPOAE screening test results Basic results view Please note that the initial result screen depends on the module settings in "DPOAE Preferences" If the required criteria were met, the result will be PASS. If not, it will be REFER.
  • Page 6 Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE Audiometry module Connect the headphones to the blue connector and select the Audiometry: Screening button from the start menu. Select the ear on which you want to start the test. Audiometry test screen Select the stimulus levels Stimulus on indicator...
  • Page 7 Bio-logic AuDX PRO FLEX ® ® QUICK START GUIDE As you present the stimulus, the current level will be displayed on the corresponding Tapping twice on a (yellow) square. Once square will display NR the hearing status for (no response). a frequency has been Once all frequencies have been tested for one ear, you can identified, move on...
  • Page 8: Copyright Notice

    Internet: www.pathme.de Healthcare solutions with one thing in mind. You. ©2020 Natus Medical Incorporated. All Rights Reserved. All product names appearing on this document are trademarks or registered trademarks owned, licensed to, promoted or distributed by Natus Medical Incorporated, its subsidiaries or affiliates. 7-26-1373-EN Rev01 Natus Medical Incorporated natus.com...