Technical Data; Heatable Sample Gas Line Cgwb 13 - ABB CGWB 13 Operator's Manual

Heatable sample gas lines
Table of Contents

Advertisement

C
4
H A P T E R

Technical Data

Heatable Sample Gas Line CGWB 13

Line with frost protection
Outer covering
PVC
Sample gas hose
PTFE
Line end assembly
various options
(see specification sheet)
Heating
self-regulating strip heater
Power supply
230 V or 115 V, 50...60 Hz
Heating power
16 W/m
Current consumption Making/hold current
= 65/25 mA
Holding temperature
+20...+50 °C
Ambient temperature –30...+60 °C
Outer diameter
30 mm
Sample gas hose
6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm
Maximum line length 130 m
Line bushing
M42
Minimum bending
300 mm
radius
Weight
approx. 1 kg/m
Connecting cable for
3 x 2.5 mm
heating (permanently
connected)
Connecting cable for
Pt100 (permanently
connected)
Self-regulating strip heater
PVC
PTFE
various options
(see specification sheet)
self-regulating strip heater
230 V or 115 V, 50...60 Hz
30 W/m
0.15 A/m
+90...+110 °C
–30...+60 °C
43 mm
90 m
M42
300 mm
approx. 1 kg/m
; length 5 m
3 x 2.5 mm
2
Parallel strip heater,
regulated
PVC
PTFE
various options
(see specification sheet)
CPD heating element,
regulated
230 V or 115 V, 50...60 Hz
60 W/m
0.3 A/m
200 °C
–30...+60 °C
40 mm
60 m
M42
300 mm
approx. 1 kg/m
; length 5 m
3 x 2.5 mm
2
LIYCI 2 x 0.14 mm
length 5 m
27
; length 5 m
2
;
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tbl 01-sTbl 01-cTbl 01-e