IV.
Adjustment of incubator:
1. <SET>
A: Adjustment of temperature
Press SET and <+> or <-> for adjusting of temperature. For a quicker setting press <+>
or <-> for 3 seconds. Press SET again to quit the menu.
B: Adjustment of menu
Press SET for 3 seconds to enter the system menu. Press <+> or <-> to select the
required menu setting. Press again <+> or <-> to enter the submenu. Press <+> or <->
again to set the parameters. Finally, press SET/RST or wait for 5 seconds for he system
to close automatically.
2. <+>: Increasing menu parameters
3. <->: Decreasing menu parameters
Operation:
Operating status:
As soon as the device is switched on, the red "working lamp" on the display´s left side is
glowing. LED indicates the actual temperature inside the incubator.
Instruction for function:
 (HU+HD) heating element
HU and HD are factory-set and should not be modified by the user
himself.
 (CA) calibration of temperature sensor
This Value is factory-set to 0°K. In case that the temperature is measured incorrectly
(test: place the thermometer for approx. 1 hour inside the closed breeding machine
and compare it with the displayed temperature) you can adjust the temperature
difference by the following means:
1.
2.
3.
4.
5.
(HS+LS) Adjust the limit of the upper/ lower temperature
The upper / lower limit of temperature setting can be restricted
1.
2.
3.
4.
5.
Adjust lower limit
1.
2.
3.
4.
© by WilTec Wildanger Technik GmbH
http://www.WilTec.de
http://www.aoyue.eu
http://www.teichtip.de
Adjust temperature sensor deviation to actual temperature
Press SET for 3 sec.
Press + or + until CA is displayed
Press SET again
By + and - you modify the parameter for the temperature difference
Temperature SET adjust upper limit
Press SET for 3 sec.
Press + or – until HS is displayed
Press SET again
By + and – you modify the parameter fort he temperature difference
Press SET for 3 sec.
Press + or – until LS is displayed
Press SET again
By + and – you modify the parameter for the temperature difference. If there is
not pressed any button within 5 sec., the last value is saved and the manu is
quit.