Rocket Medical CRAFT R29655 Service Manual

Rocket Medical CRAFT R29655 Service Manual

Oocyte aspiration pumps
Hide thumbs Also See for CRAFT R29655:

Advertisement

Quick Links

SERVICE MANUAL
®
Rocket
CRAFT™ and
DUOVAC™
Oocyte Aspiration Pumps

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CRAFT R29655 and is the answer not in the manual?

Questions and answers

Dan Gehlbach, MD
August 14, 2025

How do we calibrate the pump based on the lumen size of the retrieval needle and length of the tubing. We use a 17 gauge single lumen needle and tubing that is 250 mm long.

Summary of Contents for Rocket Medical CRAFT R29655

  • Page 1 SERVICE MANUAL ® Rocket CRAFT™ and DUOVAC™ Oocyte Aspiration Pumps...
  • Page 2: Table Of Contents

    TECHNICAL SPECIFICATIONS ........................28 CLASSIFICATION ............................ 28 15.1. SPECIFICATIONS ............................ 28 15.2. DRAWINGS..............................29 GENERAL ASSEMBLY..........................29 16.1. GENERAL ELECTRICAL ........................... 30 16.2. GENERAL INSTALLATION ........................31 16.3. Rev:11 Copyright© 2012-20 Rocket Medical plc. . All rights reserved. ZDOCK226 2020-11-16...
  • Page 3: General Assembly

    5. Water Trap Connection Port – recommended for use with R57685 Rocket Craft Pump Water Trap Sets 6. Disposable Water Trap Set 7. Dual Footswitch: BLACK: Standard: 50-250mmHg and WHITE: High Vacuum 440mmHg footswitch Rev:11 Copyright© 2012-20 Rocket Medical plc. . All rights reserved. ZDOCK226 2020-11-16...
  • Page 4: Rocket Craft™ Duo-Vac Suction Pump - Internal View

    3. Motor 220VAC 50Hz OR 110VAC 60Hz – model dependent 4. Vacuum Outlet Port 5. Pump Head including Valve Housing 6. Internal Silencer with Filter Chamber 7. Mains Power Input Rev:11 Copyright© 2012-20 Rocket Medical plc. . All rights reserved. ZDOCK226 2020-11-16...
  • Page 5: Rocket Craft™ Suction Pump

    R57685 Rocket Craft Pump Water Trap Sets 6. Disposable Water Trap Set 7. Footswitch Connection Port 8. Medium Vacuum (Standard) 50-250mmHg footswitch Footswitch supplied on Rocket Craft Suction Pumps from 2013 onwards Rev:11 Copyright© 2012-20 Rocket Medical plc. . All rights reserved. ZDOCK226 2020-11-16...
  • Page 6: Rocket Craft™ Suction Pump - Internal View

    3. Motor 220VAC 50Hz OR 110VAC 60Hz – model dependent 4. Vacuum Outlet Port 5. Pump Head including Valve Housing 6. Internal Silencer with Filter Chamber 7. Main Power Input Rev:11 Copyright© 2012-20 Rocket Medical plc. . All rights reserved. ZDOCK226 2020-11-16...
  • Page 7: General Description

    Craft I, (1984) ‘Clinical Methodology’ British Journal of Hospital Medicine 90-102 Reeves G, Scott R T, et al (1989) Journal of Assisted Reproduction and Genetics Volume 6, Number 6 / December, 1989 Rev:11 Copyright© 2012-20 Rocket Medical plc. . All rights reserved. ZDOCK226 2020-11-16...
  • Page 8: General Information

    This manual contains information that is subject to copyright. All rights reserved. WARNING: READ THIS This manual should not be photocopied, duplicated or distributed completely, or in part, without MANUAL the written approval of Rocket Medical plc. CAREFULLY: Please familiarise yourself with the contents of this manual before 3.2.
  • Page 9: Service Agents

    3.5. SERVICE AGENTS Rocket Craft™ Pumps typically require annual maintenance; however, it is recommended that they are serviced and calibrated annually at an approved Rocket Medical plc service facility. Failure to service the pump at the indicated intervals may invalidate the Warranty.
  • Page 10: Supply Voltage Selection

    Do not expose to direct sunlight. Do not expose to flammable gases. Operating temperature range: +5°C to +35°C WARNING: Device can cause explosion in the presence of flammable gases. Rev:11 Copyright© 2012-20 Rocket Medical plc. . All rights reserved. ZDOCK226 2020-11-16...
  • Page 11: Symbols Used On Craft Suction And Duo-Vac Pumps

    Read the Instruction for Use before connection and use Device is for Single Use Only Device is sterilised by Ethylene Oxide This device is not manufactured with natural rubber latex CE Mark Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 12: Annual Service

    Rocket CRAFT™ Oocyte Aspiration Pumps – Service Manual 4. ANNUAL SERVICE Service Kits: a. Filter Element: b. Pump Service Kit: Tools Required: a. Pozidriv 1 Screwdriver b. ½” Spanner c. 7mm Spanner Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 13: Replacing The Filter Element

    4. Locate the filter housing (outlined in black in the right-hand image). Note that the location differs in the Craft Suction Unit. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 14 7. Replace the top of the filter in the housing, ensuring that the lid seats securely. Take care to avoid trapping or pinching the ‘o’-ring. IMPORTANT Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 15: Replacing The Pump Valve And Diaphragm Set

    Undo and remove the centre screw to release the head assembly. Remove the pipe from the inlet and separate the two head halves with a gentle twisting motion. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 16 Maintaining the valve and head orientation, refit the head upper. When the head is assembled, refit and tighten the screw to secure the head assembly. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 17 11. Refit the new diaphragm, reusing the old washer and screw. Ensure that the outer holes align with the crankcase. DO NOT OVERTIGHTEN THE SCREW IMPORTANT Refit the spacer if fitted originally. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 18 DO NOT OVERTIGHTEN THE SCREWS Note the head alignment IMPORTANT – connect the tube to the inlet. 14. Refit the pump silencer to the head outlet. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 19: Repairs And Replacements

    Loosen and remove the securing nut. Replace the connector and tighten the securing nut. Re-fit the tubing by pushing over the connector barb. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 20: Replacing The Gauge In Duo-Vac And Suction Models

    6. Replace the gauge, reconnect tubing and insert through the panel. 7. Replace the gauge mounting plate, refit the sprung washers and tighten the wingnuts. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 21: Bottle Bracket Replacement Using Service Kit B02-190S

    Ensure that the earth lead is reattached where fitted. When fitting the bracket, ensure that the earth lead is secured to the stud using the arrangement shown below. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 22: Final Testing

    13 Repeat from #6 for each specified accuracy of the test values 200mmHg tolerance before indicated in the table returning the pump to (right). clinical service. Specified accuracy: ±10% of test value Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 23: Fuse Replacement

    2. Using a small screwdriver, carefully ease out the fuse carrier to reveal the fuses 3. To ensure continued protection REPLACE WITH FUSES OF IDENTICAL RATING IMPORTANT Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 24: Troubleshooting

    Check all the wiring plugs are connected (see section 2.2). • Check vacuum control dial is fully open. • Check gauge when Vacuum Control dial is fully closed – needle should read at least 400mmHg. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 25: Cleaning The Pump Casing

    REPAIR OR SERVICE Except under warranty, all service and/or repair should be organized via an authorized service provider. For warranty repairs, please contact Rocket Medical customer services. HAZARD: All devices to be returned must be prepared as described below for the protection of the Do NOT include used servicing team and for safety during transport.
  • Page 26: Storage

    0.51Kg – DUOVAC (metal) 0.38kg – Craft Suction (metal) Protect the device from ingress of liquid. Should any liquid enter the device, discontinue use immediately and refer to an authorised service agent Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20...
  • Page 27: Warranty

    Such warranties are extended only with respect to the purchase of the Products directly from Rocket Medical plc. as new merchandise and are extended to the first Buyer thereof, other than for resale.
  • Page 28: Technical Specifications

    Unit – 1.9Kg • Foot Switch 0.51Kg Vacuum Ranges: • Medium vacuum: -20mmHg to -200mmHg in 20mmHg increments • High vacuum: -440mmHg Vacuum Range Accuracy: ±10% of value Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 29: Drawings

    Rocket CRAFT™ Oocyte Aspiration Pumps – Service Manual 16. DRAWINGS GENERAL ASSEMBLY 16.1. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 30: General Electrical

    Rocket CRAFT™ Oocyte Aspiration Pumps – Service Manual GENERAL ELECTRICAL 16.2. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...
  • Page 31: General Installation

    Rocket CRAFT™ Oocyte Aspiration Pumps – Service Manual GENERAL INSTALLATION 16.3. Rocket Medical plc. . All rights reserved. ZDOCK226 Copyright© 2012-20 Rev:11 2020-11-16...

This manual is also suitable for:

Duovac r29660Craft r29654Craft duo-vac r29660Craft duo-vac r29661CraftCraft duo-vac ... Show all

Table of Contents