Ambient Temperature Range
Measuring device: -40 to +80 °C (-40 to +176 °F)
Note
It may not be possible to use the bluetooth connection at ambient temperatures
> 60 °C (140 °F).
Operating the Device Outdoors in Strong Sunlight
1.
Mount the device in the shade.
2.
Avoid direct sunlight, particularly in warm climatic regions.
3.
Use a weather protection cover.
11.2
Storage Temperature
-40 to +80 °C (-40 to +176 °F)
11.3
Climate Class
DIN EN 60068-2-38 (test Z/AD)
11.4
Operating Altitude according to IEC 61010-1 Ed.3
Generally up to 2000 m (6600 foot) above sea level.
11.5
Degree of Protection
Tested according to:
•
IP66, NEMA 4X
•
IP68, NEMA 6P (24 h at 1.83 m (6.00 foot) under water)
11.6
Vibration Resistance
DIN EN 60068-2-64/IEC 60068-2-64: 20 to 2 000 Hz, 1 (m/s
11.7
Electromagnetic Compatibility (EMC)
Electromagnetic compatibility in accordance with all of the relevant requirements outlined
in the EN 61000 series and NAMUR Recommendation EMC (NE 21). Details are provided
in the Declaration of Conformity, refer to www.pepperl+fuchs.de.