Advertisement

2
Site Requirements and Specifications

Site Requirements

Site Requirements
A suitable environment is important to ensure optimal performance of the
instrument.
Power Considerations
The module power supply has wide ranging capability. It accepts any line
voltage in the range described in
is no voltage selector in the rear of the module. There are also no
externally accessible fuses, because automatic electronic fuses are
implemented in the power supply.
Hazard of electrical shock or damage of your instrumentation
WA R N I N G
can result, if the devices are connected to a line voltage higher than specified.
The module is partially energized when switched off, as long as the power cord is
WA R N I N G
plugged in.
Repair work at the module can lead to personal injuries, e.g. electrical shock, when
the cover is opened and the module is connected to power.
Inaccessible power plug.
C A U T I O N
In case of emergency it must be possible to disconnect the instrument from the power
line at any time.
22
Connect your instrument to the specified line voltage only.
Always unplug the power cable before opening the cover.
Do not connect the power cable to the instrument while the covers are removed.
Make sure the power connector of the instrument can be easily reached and
unplugged.
Provide sufficient space behind the power socket of the instrument to unplug the
cable.
Table 2
on page 25. Consequently there
Agilent 1260 Infinity TCC User Manual