Lamtec LT3 User Manual
Lamtec LT3 User Manual

Lamtec LT3 User Manual

Lambda transmitter, lambda probe
Hide thumbs Also See for LT3:
Table of Contents

Advertisement

User Manual
Lambda Transmitter LT3
Lambda Probe LS2
Sensors and Systems for Combustion Engineering
www.lamtec.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LT3 and is the answer not in the manual?

Questions and answers

Summary of Contents for Lamtec LT3

  • Page 1 User Manual Lambda Transmitter LT3 Lambda Probe LS2 Sensors and Systems for Combustion Engineering www.lamtec.de...
  • Page 3: Table Of Contents

    Table of Contents Table of Contents Important Information about the Manual ........4 Validity of these Instructions .
  • Page 4 Table of Contents 6.1.1 Faults ............37 6.1.2 Warnings .
  • Page 5: Important Information About The Manual

    Important Information about the Manual Validity of these Instructions This manual describe the Lambda Transmitter LT3 with all required components. The informa- tion in this document applies to the software version V1.09.0.0. If you utilise a different version, this can lead to other effects to your device than those described in this manual.
  • Page 6: General Safety Instructions

    General Safety Instructions General Safety Instructions Classification of the Safety Instructions and Warnings The following symbols are used in this document to draw the user's attention to important safe- ty information. They are located at points where the information is required. It is essential that the safety information is observed and followed, and that applies particularly to the warnings.
  • Page 7: Proper Use - Conditions Of Use

    Auxiliary devices or those from other manufacturers must be rec- ommended or authorised by LAMTEC. Provided the safety information and operating direc- tives specified in this manual are observed, this device should not present any risks in terms...
  • Page 8: Permissible Users

    User groups For the handling of the Lambda Transmitter LT3, three user groups are required: • Service technicians of LAMTEC or its OEM customers and/or trained customer personnel: Qualified technicians/engineers ...
  • Page 9: Safety Equipment/Safety Measures

    To avoid damage, always observe the respective security notices. Preventive measures for improving operating safety If the LT3 is used in conjunction with control and monitoring technology, the operator must en- sure that any breakdown or failure of the LT3 device does not cause inadmissible damage or dangerous operating states.
  • Page 10 General Safety Instructions NOTICE Do not switch off the Lambda Transmitter LT3 as long as the LS2 Lambda Probe is mounted. Not even if the associated plant has been brought to a standstill. Residual gases will cause corrosion and could damage the probe.
  • Page 11: Product Description

    Product Description Product Description Design and Accessory LS2 Lambda Probe The Lambda Probe LS2 provide the in situ (direct in the flue gas) measurement of O concen- tration in exhaust gases in the super-stoichiometric range (>1). Fig. 3-1 Lambda Probe LS2 standard design Lambda Probe LS2 in standard housing up to 300 °C (572 °F) Default cable length is 2 m (6.5 ft), FEP, with connecting plug Probe installation fitting (PIF)
  • Page 12 Product Description Fig. 3-3 Lambda ProbeLS2-HT for manual purging (for application in high dusty environments) Lambda Probe LS2-HT up to 300 °C (572 °F) Temperature at probe heat max.100 °C (212 °F) Default cable length is 2 m (6.5 ft), FEP, with connecting plug High dust reception flange with pneumatic connections...
  • Page 13: Display And Operational Controls

    Display and Operational Controls Display and Operational Controls To display and operate LT3, the user interface is integrated into the front door of LT3 (included in the standard). Functions: • Reading of the O measured values • Password entry Information  regarding the probe, fuel, warnings, faults, software version, •...
  • Page 14: Operation

    Display and Operational Controls Operation Using the arrow keys you can navigate within the menus. The keys mov their selection to the right or the left hand. ENTER ENTER takes you to menu entries or to the edit mode in the case of selected pa- rameters or values.
  • Page 15: Information Menu Structure

    Display and Operational Controls 4.2.2 Information Menu Structure 4.2.3 Calibration Menu Structure Enter the password for access level 1 (see chapter 5.1.4 Enter the Password for the Access Level).
  • Page 16: Settings Menu Structure

    Display and Operational Controls 4.2.4 Settings Menu Structure Enter the password for access level 1 (see chapter 5.1.4 Enter the Password for the Access Level).
  • Page 17: Status Line

    Display and Operational Controls Status Line The status line contains information about LT3, including the device family. Fig. 4-3 UI300-LT-V2 front view status line 1. Status line Description status line: Active fault/warning Operating mode OK Heat-up (cold start) Status of limit values 1-4...
  • Page 18: Main Menu

    Display and Operational Controls Main Menu Basic display Main menu The display shows the O value, resolution 0,1 % Press ENTER to access the main menu. Meaning of the symbols: Password entry Information regarding the probe and fuel, warnings and faults, software version, CRC, and serial number Calibration of measurement Settings (maintenance, filter time, analogue output, probe replacement, display,...
  • Page 19: Main Menu - Information

    4.4.1 Main Menu - Information Meaning of the symbols: Measured values and probe data Fault/warning history Software version of LT3 and display CRC checksums Serial number Manual reset of the limit values if they are set to manual reset 4.4.2 Main Menu - Calibration Enter the password for access level 1 (see chapter 5.1.4 Enter the Password for the Access...
  • Page 20: Main Menu - Settings

    Display and Operational Controls 4.4.3 Main Menu - Settings Enter the password for access level 1 (see chapter 5.1.4 Enter the Password for the Access Level). Meaning of the Symbols Required Access Level Maintenance mode none Filter time of the measurement values none Analogue outputs 1 (customer) or higher...
  • Page 21: Maintenance

    Maintenance Maintenance As a result of extensive self-diagnosis, the measuring system is essentially maintenance-free. Maintenance is restricted to the calibration and the cleaning of dust or other deposits; as well as the cyclic replacement of the ZrO measuring element. Probe Testing and Calibration The checking / calibration has to be built (advantageously installed) in at normal operating temperature and carried out under operating conditions.
  • Page 22 Maintenance In MEASURED VALUES and PROBE DATA, select: Display of the O measured values, Display of the R internal probe resistances Display of the current fuel selection The O probe voltage U-O stabilises to val- ues between +10 ... -20 mV. Die R internal probe resistance lies between 15 ...
  • Page 23: Enter The Password For The Access Level

    Maintenance 5.1.4 Enter the Password for the Access Level In the main PASSWORD ENTRY menu, select Enter password via ENTER Access level 1: Customer Level Factory setting: 0000 – access to customer functions only Customers can change the password for access level 1. For details ask the burner manufacturer/supplier Access level 2: Service level...
  • Page 24: Activate/Deactivate Maintenance Mode

    Menu 460 ACTIVATE/DEACTIVATE MAIN- TENANCE MODE opens. Use the keys and ENTER activate/deactivate the MAINTENANCE mode. Activate maintenance mode Deactivate maintenance mode If both parameters (UI300-LT-V2 and LT3) are the same confirm with ENTER Respect the time limit of 8 s...
  • Page 25: Response Of The Internal Resistance Regulation

    Maintenance Wait until the display shows the transmis- sion’s end. Return to the SETTINGS menu with the key If the procedure is cancelled or in case of timeout this display is shown. BACK the SETTINGS menu. • The status line shows this symbol 5.1.6 Response of the Internal Resistance Regulation The internal ceramic resistance between the reference and O...
  • Page 26: Premature Cold Start Termination

    Maintenance 5.1.7 Premature Cold Start Termination Occasionally, it might be necessary to terminate the COLD START prematurely, e.g., after a short-term power failure, but not before a minimum waiting period of 120 seconds has expired. NOTICE A premature cold start termination leads directly into measurement operation without an addi- tional prompt.
  • Page 27: Calibrating The Probe

    Maintenance 5.1.8 Calibrating the Probe Maintenance mode has to be activated before running a calibration process, see chapter 5.1.5 Activate/deactivate Maintenance Mode. This deactivates probe failures and avoids fault con- ditions. After finishing the calibration process successfully, maintenance mode must be deac- tivated.
  • Page 28: 5.1.8.1 Offset Calibration Perform To 21 % O

    Maintenance 5.1.8.1 Offset Calibration perform to 21 % O value is equivalent to 15 (see menu 211 and Before offset calibration, check whether R 213 in chapter 5.1.3 Reading the Measured Values). Otherwise, proceed as described in chapter 5.1.6 Response of the Internal Resistance Reg- ulation.
  • Page 29 Maintenance Using the keys and ENTER triggers the offset calibration Confirm the setting with ENTER within 10 s, otherwise the value is rejected. The U-O value stabilises to values between +10 ... -20 mV. The R internal probe resistance value is 15 .
  • Page 30: O 2 Calibrating In The Operating Point With Reference Measurement

    Maintenance 5.1.8.2 Calibrating in the Operating Point with Reference Measurement • Activate MAINTENANCE mode. NOTICE At least access level 1 required! The calibration can be performed by reference measurement using an external measuring de- vice or by using test gas at a known O concentration (LS2default design only).
  • Page 31: Checking/Calibrating With Test Gas

    Maintenance Adjust and calibrate the O values with Confirm the calibration within 5 s with Otherwise the calibration will be rejected. • Deactivate MAINTENANCE mode. 5.1.9 Checking/Calibrating with Test Gas For the standard design of the LS2 Lambda Probe, a special testing device is required (type 650R1015).
  • Page 32 Maintenance Fig. 5-2 Testing device laying on the table Probe LS2, KS1, KS1D in Standard housing Hose Particle filter Test gas connection ON Plug in nozzle for flue gas channel Test gas outlet Hose Fig. 5-3 Testing device plugged into flue gas channel...
  • Page 33 Maintenance Test probe 1. Install probe: Push probe into the testing device until stop. 2. Heat up probe: A cold probe has to be heated up for at least 45 minutes in the installed condition. Should an already heated up probe be installed in the testing device, then wait for 15 minutes. CAUTION! The testing device with plugged in heated probe would get hot.
  • Page 34: Checking/Calibrating With Reference Measurement

    Maintenance Table 4:Test gas matrix for testing or calibrating the respective probe Functional check/calibration Probe with test gas 1  Offset 2  O Table 5:Test gas composition Composition Test gas [Vol.%] [Vol.%] Rest Rest Recommendation: To check the O electrode, use a test gas with 3 vol.
  • Page 35: Settings

    Maintenance 5.1.12 Settings Adjusting the settings depends on the current access level. Access level 2 (service level) al- lows to adjust all settings. Select in the MAIN MENU for the set- tings. SETTINGS allows the following: Maintenance mode (level 0) see chapter 5.1.12.1 Maintenance Mode Filter time for the measured values (level 0)
  • Page 36: 5.1.12.4 Display

    Maintenance 5.1.12.4 Display Display (no access level is required): Brightness Contrast Time in seconds until the background lighting is shut OFF Switch from °C to °F and vice versa...
  • Page 37: Limit Values

    Maintenance 5.1.12.5 Limit Values Limit values Click to proceed. Use the keys to change the limit values. Limit value = OFF Monitoring exceeding Monitoring undershooting Setting of the limit value Adjusting the trigger time of the limit value NO - Limit value is resets automatically YES - Manual acknowledgment of limit value is required (see chapter 4.4.1 Main Menu - Information) If a limit value is undershot/exceeded, this is indicated in the status line by a down/up arrow...
  • Page 38: Correcting Faults

    10 s Help: - Check the wiring. LT3 Term. 13 and 14  disconnect the probe - probe heating at LT3 or in and measure the heater resistance between wire pos. 13 and 14 on the probe side. Resistance approx. 9 ... 10  in heated state.
  • Page 39 OFF current too high Heating control circuit open Heater controller at limit stop for longer than 30 seconds. Help: Check wiring between probe and LT3 (check probe signals terminals 10/ Temperature of the O sensor too high (R too low) is under 10 ...
  • Page 40 Look for a new installation position. - Flow speed too high. - Heating defective and probe cold. - Sensor breakage. - Check wiring between probe and LT3 (signal terminals 10/11/12) under 10 , probe is too hot, internal resistance regulation is not If R active, or incorrect setpoint value.
  • Page 41: Warnings

    Trigger value of current internal resistance (x0.1 ), /0 - 65535 tolerance time of 3 seconds. 016 (HP) Internal warning 116 (ÜP) Serial number, customer ID or service password reset. Contact the manufacturer. Display if customer ID in LT3 and UI do not correspond.
  • Page 42: Call Of The Fault History

    Correcting Faults 6.1.3 Call of the Fault History In the MAIN MENU, select for informa- tion. In INFORMATION, select for the fault history. Fault history Number of the entry 001 Fault number of the main processor If the fault number is displayed as "101" (+100), there is a fault in the monitoring processor.
  • Page 43: Decommissioning

    NOTICE Do not switch off the LT3 Lambda Transmitter as long as the LS2 Lambda Probe is mounted. Not even if the associated plant has been brought to a standstill. Residual gases will cause corrosion and could damage system components.
  • Page 44: Options

    Options Options Analogue Outputs via LSB Module Current, alternative Voltage, LSB address 19 8.1.1 Functional Description • Current module: 4 analogue outputs 0/4 ... 20 mA • Voltage module: 4 analogue outputs 0/2 ... 10 VDC • Possible to quickly wire several modules by means of strapping plugs The LSB modules are universally applicable output modules, which are controlled via the LSB SYSTEM BUS.
  • Page 45: Factory Setting Of Analogue Outputs Via Lsb Module

    Analogue output 4 24 VDC Voltage supply for LT3 Terminals 77-/78+ CAN H/L LAMTEC SYSTEM BUS for LT3 Terminals 74 H/75 L 8.1.2 Factory Setting of Analogue Outputs via LSB Module The module is activated at the factory. In case of a later installation or exchange, the LSB ad- dress 19 must be set on the module using only 2 rotary switches.
  • Page 46: Conversion Of The Output Range Via The User Interface

    Options 8.1.3 Conversion of the Output Range via the User Interface Access level 1 or higher required. Select in the main menu to switch to the settings menu. Select to switch to the analogue outputs. The analogue outputs menu provides the fol- lowing setting options: Current range or voltage range Output range...
  • Page 47: Digital Outputs Via Lsb Module, Lsb Address 03 And 51

    Rotary switch for setting the ones 33/34 Relay output 3 LSB address LSB Bus Connector Plug 43/44 Relay output 4 Manual activation 24 VDC Voltage supply for LT3 Termi- nals 77-/78+ CAN H/L LAMTEC SYSTEM BUS for LT3 Terminals 74 H/75 L...
  • Page 48: Factory Setting Of The Digital Outputs

    Options 8.2.2 Factory Setting of the Digital Outputs The LSB module for digital outputs 1 ... 4 is activated at the factory. In case of a later installa- tion or exchange, the LSB address 03 must be set on the module using only 2 rotary switches. Digital output 1 Faults Term.
  • Page 49: Digital Inputs Via The Lsb Module, Lsb Address 11 And 55

    Digital input 3 LSB address Rotary switch for setting the ones 4+/4- Digital input 4 LSB address Jumper plug 24 VDC Power supply for LT3 Terminals 77-/78+ Manual activation CAN H/L LAMTEC SYSTEM BUS for LT3 Terminals 74 H/75 L...
  • Page 50: Factory Settings Of The Digital Inputs

    Options 8.3.2 Factory Settings of the Digital Inputs Digital input 1 Trigger OFFSET CALIBRATION Digital input 2 RESET FAULT Digital input 4 DEACTIVATION LIMIT VALUE 1 ... 4 Digital input 5 RESET LIMIT VALUE 1 ... 4 Digital input 8 DEACTIVATION CALIBRATION NOTICE Manual activation of the relay outputs:...
  • Page 51: Lsb Module For Calculating Combustion Efficiency

    Options LSB Module for Calculating Combustion Efficiency Setting of LSB address not necessary. 8.4.1 Functional Description Properties: • Two Pt100 temperature inputs to record the flue gas temperature and inlet temperature • Two analogue outputs 0/4 ... 20 mA to emit the flue gas temperature and efficiency •...
  • Page 52 Options NOTICE The termination resistor (120 ) must be installed and activated on the first and the last BUS device. Avoid stub line! The terminal resistance can be set via DIP switch 1. Terminal assignment: 10 / 11 / 12 Pt100 temperature input for recording the flue gas temperature 0 ... 400 °C/ 32 °F ...
  • Page 53: Storage

    Storage Storage Storage Conditions NOTICE  Do not store the device outside without any protection!  When uninstalling, protect the ends of cables and the connectors from corrosion and soil- ing. Corroded connectors can cause malfunctions.  Always store in a dry location and, if possible, keep in the original packaging. ...
  • Page 54: Disposal Notes

    Disposal Notes Disposal Notes 10.1 Environmental Protection, Waste Disposal The design of Lambda Transmitter and LS2 is also based on environmental considerations. The modules can easily be separated and sorted into distinct types, and recycled accordingly. The device contains electrical and electronic components and must not be disposed of as do- mestic waste.
  • Page 55: Appendix

    Wet/Dry Measurement Deviations, Conversion Table NOTICE The LT3 carries out measurements directly in the humid flue gases (wet measurement). When extractive devices are used, flue gases are removed and prepared. Dry measurements are normally used here, since the humidity has been extracted from the flue gas.
  • Page 56: Eu Declaration Of Conformity

    Appendix 11.2 EU Declaration of Conformity...
  • Page 57 Appendix...
  • Page 58 The information in this publication is subject to technical changes. LAMTEC Meß- und Regeltechnik für Feuerungen GmbH & Co. KG Josef-Reiert-Straße 26 D-69190 Walldorf Telefon: +49 (0) 6227 6052-0 info@lamtec.de Telefax: +49 (0) 6227 6052-57 www.lamtec.de Printed in Germany | Copyright  2021...

This manual is also suitable for:

Ls2Ks1d

Table of Contents

Save PDF