Boot and error messages of the operating PC

Operating system errors

"Start-Up-Message: No HOST-Communication"
Meaning:
"Interface-Error: DUEP-POLLING-TIME-OUT"
Meaning:
"Interface-Error: HOST-DATA-OVERFLOW"
Meaning:
"Interface-Error: NEW-INIT-REQUEST"
Meaning:
"Interface-Error: TIME-OUT-INIT"
Meaning:
"Interface-Error: TIME-OUT-SEND"
Meaning:
"Interface-Error: INVALID-DPR-FLAG"
Meaning:
"Interface-Error: CHECKSUM-ERROR"
Meaning:
"Interface-Error: RECEIVE-BUFFER FULL"
Meaning:
"Interface-Error: SEND-BUFFER FULL"
Meaning:
Siemens AG
Medical Solutions
This message is displayed during booting of the operating PC as long as no
Dual Port RAM (DPR) communication is running;
The plug-in card reports via the DPR interface that polling is no longer rec-
ognized at the DUEP interface
(Communication fault);
If the HOST transmits data faster than can be forwarded to the operating PC,
then this message is displayed;
The plug-in card requests resynchronization of the DPR communication
because of internal errors (e.g. watchdog reset);
A timeout was detected during the booting synchronization of the DPR inter-
face;
A timeout has occurred while transmitting data via DPR interface to the plug-
in card (DPR communication error);
The DPR communication protocol was not observed
(DPR communication fault);
A checksum error has been detected while receiving data via DPR interface
(DPR communication error);
The operating PC has received data too fast and an internal data overflow
has occurred;
The operating PC has generated more data blocks than can be sent and
thus triggered an internal data overflow;