W Drive Am3; O Drive Am4 (C-Arm); Board Errors, General - Siemens LITHOSTAR Multiline Troubleshooting Instructions

Hide thumbs Also See for LITHOSTAR Multiline:
Table of Contents

Advertisement

1 - 2

W drive am3

Error number
Meaning:
30 LM_628_error
31 Control_error
32 Position accuracy not reached in stated stabilizing time
33 Motor not initialized, drive command received
34 Motor moves without actuation
35 Motor does not move when actuated
36 Speed not reached or exceeded
37 Motor moves in wrong direction
38 Motor unsuccessfully actuated three times
39 Faulty initialization of the ASIC

O drive am4 (C-Arm)

Error number:
Meaning:
40 LM_628_error
41 Control_error
42 Position accuracy not reached in stated stabilizing time
43 Motor not initialized, drive command received
44 Motor moves without actuation
45 Motor does not move when actuated
46 Speed not reached or exceeded
47 Motor moves in wrong direction
48 Motor unsuccessfully actuated three times
49 Faulty initialization of the ASIC

Board errors, general

Error number:
Meaning:
60 No first contact with HOST
61 TIMEOUT, data exchange disrupted
62 CHECKSUM_error
LITHOSTAR Multiline
System
RXL2-120.072.03
Rev. 04
06.02
Motor controller D4a
Page 2 of 2
CS PS 24
1
1
1
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

Table of Contents