Aquatec® Ocean E-VIP
Malfunction
Battery does not charge
Implementing Reset
1. Hold down the following three buttons simultaneously for about three seconds:
Emergency stop button E,
Forward button C and
Up button H.
An acoustic signal is heard while the three buttons are held down.
2. When the service indicator lamp F turns off, release the buttons. The device is ready to use again.
3. Then move the drives for adjusting the height to the lowest position until you hear the drives switch off.
4. Move the tilt drive to the rear end position until you hear the drive switch off. (The red LED flashes and an acoustic
signal is heard).
If the fault recurs after the reset, replace the device or the faulty parts and send them to your distributor (addresses
see last page).
32
Possible Cause(s)
Hand control faulty
Faulty battery
Connector not correctly inserted
Remedy
Implement reset (see below table)
or
Replace hand control
Replace battery
Check that all connectors are positioned
correctly