3.4.2.2.6 Unscrew the bolts in the back bumpers and
remove the bumpers.
Back
bumpers
3.4.2.2.7 Lift the top cover off of the table to reveal the
base actuator attachment bolts.
3.4.2.2.8 Brace the table by placing a block under the
frame close to the actuator being replaced.
Actuator
Bolts
3.4.2.2.9 Unscrew the bolts on the top of the actuator
and remove the bolts. Then slide the actuator
out from underneath the table and allow the
table to rest on the block.
WARNING:
The table is very heavy and must
be braced properly before removing the
actuator to ensure that the table does not fall.
3.4.2.2.10 Lay the actuator and base assembly on the
3.4.2.2.11 Install the new actuator onto the base by
3.4.2.2.12 Lift the actuator and base assembly up and
Actuator
Bolts
3.4.2.2.13 Replace the top cover and re-attach the
3.4.2.2.14 Replace the upholstered top and install the
3.4.2.2.15 Re-install the back actuator. See section 3.2.5
3.4.2.2.16 Route the wires from the base actuator back
3.4.3 If this does not work, Contact UMF Medical
Customer Service at (814)266-8726.
UMF Medical
www.umfmedical.com
side and remove the bolts from the base to
completely detach the actuator.
Actuator
Bolts
reinstalling the four bolts on the bottom of the
actuator.
CAUTION:
Ensure that the new actuator is
oriented properly before installing.
slide it back underneath the table. Install the
four bolts through the table into the top of the
actuator.
bumpers.
four bolts to secure the upholstered top.
to the Control Box and secure in place.