The Powered Back Section Is Not Moving When Activated. What Should I Do - UMF Medical 5005 Service Manual

Bariatric power treatment table
Table of Contents

Advertisement

3.1.7.4 Install new Control Box by connecting the same four
(4) screws to the table.
3.1.7.5 Install all plugs.
3.1.7.6 Connect power to table.
3.1.8 Replace Hand Control.
3.1.8.1 Disconnect power to table
3.1.8.2 Unplug Hand Control from Control Box and plug in
new Hand Control.
3.1.9 Contact UMF Customer Service at (814)266-8726.
3.2 THE POWERED BACK SECTION IS NOT MOVING WHEN
ACTIVATED. WHAT SHOULD I DO?
3.2.1 Check to make sure the table is moving up and down
properly. If not, follow instructions in section 3.1.
3.2.2 Check visible wires for damage. Inspect all wires that
are readily visible for signs of cuts, tears, or other
damage that could cause a loss of communication
between components.
3.2.3 Check plugs. Check all plugs and connections to ensure
that all wires are properly connected.
3.2.4 Swap receptacles. Plug back section actuator into
receptacle 2 on the Control Box and remove plugs from
all other actuator receptacles.
3.2.4.1 If the actuator functions properly, replace the
Control Box (see section 3.1.7).
3.2.4.2 If the actuator does not function, replace the
actuator (see section 3.2.4).
3.2.5 Replace back actuator.
3.2.5.1 Disconnect power to table.
3.2.5.2 Unplug actuator cord from Control Box and free the
cord from the table.
3.2.5.3 Remove pins on end of actuator arm and body and
remove actuator.
UMF Medical
www.umfmedical.com
Pins
6