INTRODUCTION
A
complete
procedure
standards and calibration of the Type Z Plug-In Unit is
provided in this section of the manual.
procedure are arranged in correct sequence to avoid un
necessary repetition.
The step-by-step instructions also furnish an orderly ap
proach for isolation of malfunctions which may develop.
Consequently, this procedure should be used in conjunction
with any maintenance and troubleshooting system.
If an abnormal indication is obtained at some point in
the procedure, it is not usually necessary to locate its cause
before continuing to the next step.
revealed by performing further steps will frequently simplify
the task of locating trouble. Obscure symptoms, if any,
usually show up when performing the first eleven steps of
the procedure.
For this procedure, the Z Unit should be used in con
junction with a properly calibrated Tektronix Type 540-
Series oscilloscope.
Oscilloscope
Type 541A Oscilloscope are listed in the procedure; cor
responding control settings should be used if a different
oscilloscope is used.
Test equipment used in a particular step should be left
connected at the end of that step unless instructions state
otherwise.
Similarly, controls not mentioned are assumed
to be in the positions they were in at the conclusion of the
preceding step.
If information explaining the normal operation of front-
panel controls is needed before starting the procedure,
consult the Operating Instructions section of this manual.
EQUIPMENT REQUIRED
The following equipment or its equivalent is required to
perform a complete calibration of the Type Z Plug-In Unit.
1. Square-wave generator, Tektronix Type 105 or equiva
lent.
Required specifications are:
output frequencies of approximately 1 kc and 10 kc; output
amplitude variable from 10 to 100 volts across 600-ohm
internal load.
2. Square-wave generator, Tektronix Type 107 or equiva
lent.
Required specifications are: 3 nsec risetime or less;
output frequency of approximately 1 me; output amplitude
variable to at least 0.35 volt.
for
checking
the
operational
The steps in the
Additional symptoms
control
settings for a
13-nsec risetime or less;
C A L I B R A T I O N
3. Sine-wave
generator, Tektronix
Amplitude Signal Generator or equivalent.
ifications are: frequency range of 50 kc to 13 mc; output
amplitude of 0.14 volt or more; output must be adjustable
(manually or automatically) for constant amplitude within
the above frequency range.
4. Precision
dc
voltmeter.
accuracy of 0.05% or better; resolution of 50 ^.volts or less;
must be of the nulling type with infinite impedance at null.
If a John Fluke Differential Voltmeter is available, use
Model 803 or equivalent.
5. Calibrated volt-ohm meter (VOM). Sensitivity of 20,000
Q/v at full deflection.
6.
24-J uJ uf Input Capacitance Standardize^ Tektronix Type
CS24 (Part No. 011-029) or equivalent.
7. Plug-in cable extension, Tektronix Part No. 012-038.
8. Plug-in extension, Type EP54, Tektronix Part No. 013-019.
9. 52-ohm coaxial cable, Tektronix Part No. 012-001.
10. 52-ohm 5-to-l " L" attenuator, Tektronix Part No. 011-
002
.
11. 52-ohm cable termination, Tektronix Part No. 011-001.
12. Suggested tools: screwdriver, shank 2 "
dia.;
0.050"
alien
wrench;
Tektronix Part No. 003-000; alignment tool consisting of
a handle (Tektronix Part No. 003-007), a low-capacitance
screwdriver insert with a metal bit (Tektronix Part No.
003-334, and a hexagonal core insert (Tektronix Part No.
003-310).
PRELIMINARY PROCEDURE
N O TE— M ake certain the — 150-, + 1 0 0 - , + 2 2 5 - ,
and + 350-vo lt power supplies in the oscilloscope
are regulating
and w iti
ances before attempting
Preset the Type 541A Osci
panel controls as follows:
Horizontal Display
Time/cm
Variable Time/cm
Triggering Mode
Triqqer Slope
Stability
Type
190A
Constant-
Required spec
Required
specifications
are:
long x 1/8"
low-capacitance
screwdriver,
in rated voltage toler-
to calib rate the Z Unit.
loscope (or equivalent) front-
Normal
.5 Millisec
Fully Clockwise
AC
+ Internal
Fully Clockwise
5-1