Technical Data; Eft Electrical Fast Transients Burst As Per Iec 61000-4-4 - EM TEST EFT 500 Series Manual For Operation

Eft/burst generator pulse 5/50ns
Table of Contents

Advertisement

EM TEST
4.

Technical data

4.1.

EFT Electrical Fast Transients Burst as per IEC 61000-4-4

Test Level
Open circuit
Wave shape into a 50Ω load
Rise time tr
Pulse duration td
Wave shape into a 1000Ω load
Rise time tr
Pulse duration td
Source impedance
Polarity
Trigger
Trigger of bursts
Synchronization
Burst duration td
Burst repetition rate tr
Spike frequency f
Test duration T
Output
Direct via 50Ω coaxial connector
Coupling network
DUT power mains supply
Test routines
Quick Start
Standard test as per
User test routines
Safety
Safety circuit
Design
Interfaces
RS 232
IEEE
CN port
General data
Dimensions
Weight
Power supply
Fuses
Manual for Operation
200V - 4400V ± 10%
100V – 2200V
5ns ± 30%
50ns ±30%
200V – 4400V
5ns ± 30%
35ns - 150ns
Zq = 50Ω ± 20%
positive / negative
AUTO, MANUAL, EXTERN
0° - 360°
0.1ms - 999.9ms
10ms - 9999ms
0.1kHz - 1000kHz
0:01 min - 99:59 min
To connect ext. coupling devices
To L, N, PE all combinations
AC 250 V / 16 A / 50/60 Hz
DC 250V/10A
Immediate start, all parameters adjustable during a running test
IEC 61000-4-4 level 1 ... level 5
Level X - Level Y
Customized test routines
Voltage change after T by ∆V
Frequency change after T by ∆f
Frequency sweep in one single burst
Change duration after T by ∆td
Change polarity after T
Statistical burst release
Synchronized at fixed angle
Control input ( short circuit )
as per IEC 1010, EN 61010
serial, 1200 - 19200 Baud
parallel, addresses 1-30
Control of CNE 503 for 3-ph testing
19" / 3 HUE
approx. 8kg
115V/230V +10/-15% 50/60Hz
2 x T 1A slow blow
V 2.00
Step 10V
EFT 500 Series