4-4 Repair/Replacement and Calibration
5. Remove the three #8 Phillips-head screws located on the back of the
hood, above the kick plate.
#8-32 x 0.375 L
6. Grasp the front section of the hood cover at the top and bottom center.
Pull and lift to remove while feeding the power cord through the hood
cover opening.
7. Use a long screwdriver to remove the two screws holding the hood to
the front of the siderails. Loosen the two rear screws holding the back
part of the hood to the headframe.
CE units: Remove the #10 screws that hold the hood to the uprights.
8. Gently tap the hood a half inch towards the rear until the rear screws
disengage from the key slots.
9. Spread the sides of the hood slightly apart to clear the uprights. Tilt
the hood upward, then slide it back and lift it off the treadmill.
Recommended: Vacuum the internal components before replacing the
hood or applying power. Use caution near the AC Drive Module. Do not
vacuum the drive board on the AC Drive Module.
#10 screws
MedTrack CR60 Rehabilitation Treadmill Service Manual
Push to
Remove