Removing The Treadmill Hood - Quinton MedTrack CR60 Service Manual

Rehabilitation treadmill
Hide thumbs Also See for MedTrack CR60:
Table of Contents

Advertisement

Removing the Treadmill Hood

Warning!
MedTrack CR60 Rehabilitation Treadmill Service Manual
Tachometer
Wooden blocks
Belt-tension calipers (optional)
Masking tape or heavy pencil
Although some procedures are used for several applications (removing the
hood, for example), each procedure is explained only once. Reference
procedures under their headings for subsequent applications.
All references to front, rear, left, and right are given as though you were
facing the front handrail while walking on the deck.
1. For your convenience in working, elevate the treadmill to its
maximum height, if possible.
2. Turn off the circuit breaker on the treadmill hood, then unplug the
treadmill from the power outlets.
To prevent high voltage electrical shock, wait at least two
minutes from the time you unplug the power cords before
working on or around any electrical or mechanical
component under the hood.
3. Remove the four Phillips-head screws from the configuration plate.
4. Remove the three ¼-20 Phillips-head screws located under the label
panel on the front of the hood.
Screw - #10-32 c 0.375L
Lock Washer - ext. star #10
Configuration
Plate
Hood screws
1/4-20 x 0.50L
Ext. Star Lock Washer
Repair/Replacement and Calibration 4-3